WEBVTT

00:00.000 --> 00:01.500
My name is Kwadi.

00:01.780 --> 00:05.660
I'm excited here to introduce this next panel.

00:06.180 --> 00:09.880
And the topic of which you might think is a little bit of a mystery because it's not much on the website.

00:09.880 --> 00:11.920
Who here has any idea what this panel is about?

00:12.840 --> 00:13.340
Great.

00:13.340 --> 00:14.680
And yet you still came.

00:14.680 --> 00:15.540
So that's awesome.

00:15.540 --> 00:16.660
That is amazing.

00:17.080 --> 00:21.420
I can't tell if you just need a place to sit or what, but we will proceed with some very interesting content.

00:22.020 --> 00:23.880
Primarily because this topic is very interesting, right?

00:23.880 --> 00:27.880
We're talking about hacking medical devices, hacking critical hospital infrastructure.

00:28.080 --> 00:32.720
As you can see all the great work that's being done in this village next door about keeping patients safe.

00:32.720 --> 00:35.540
And that's really the key thing we're going to talk about today.

00:35.540 --> 00:40.020
And when we say simulation, simulation comes in a lot of different flavors, has a lot of different definitions.

00:40.020 --> 00:47.800
But if we think about it, we're really talking about what could happen to patients that are connected to these devices that are vulnerable.

00:48.780 --> 00:54.780
How can we study that ahead of time because we don't want to be discovering it after the fact, right?

00:54.780 --> 01:05.440
So we have to use the critical tool of simulation, whether or not that be high-fidelity simulation when we train doctors or whether or not it's going to be simulating attacks in, in different environments so we can better understand and learn them.

01:05.700 --> 01:08.160
So that's enough of my introduction for the topic today.

01:08.160 --> 01:11.020
I'm going to go ahead and start introducing some of our panelists.

01:11.260 --> 01:12.420
Um, Dr.

01:12.460 --> 01:19.320
uh, Saxon, uh, who is listed to be on this panel had a family emergency this morning and actually has flown out of, uh, the state.

01:19.320 --> 01:20.900
So I apologize to anyone who wanted to come for Dr.

01:20.900 --> 01:23.600
Saxon, but she sends her, uh, her regards.

01:24.740 --> 01:26.260
We're going to go ahead and start off.

01:26.260 --> 01:26.880
Uh, we have Dr.

01:26.880 --> 01:28.960
Julian Goldman at the end.

01:29.140 --> 01:31.520
They're all each going to give their own introductions.

01:32.200 --> 01:32.680
We have Dr.

01:32.680 --> 01:34.380
Suzanne Schwartz.

01:35.300 --> 01:37.240
Yeah, Suzanne Schwartz!

01:38.220 --> 01:39.140
And we have Mr.

01:39.140 --> 01:40.180
David Guffrey.

01:41.500 --> 01:42.480
I'm excited.

01:47.100 --> 01:49.420
They'll now proceed with their own introductions.

01:51.060 --> 01:52.240
I'm so sorry.

01:52.240 --> 01:55.300
Listen, I went to public school and I can't spell anything.

01:55.300 --> 01:57.740
So these are, please forgive me.

01:58.520 --> 02:00.480
Yes, yeah, it's the anesthetic.

02:01.020 --> 02:01.820
Please.

02:02.680 --> 02:03.880
Of course.

02:06.400 --> 02:07.460
Thank you.

02:08.560 --> 02:10.520
Okay, so it's been given away.

02:10.520 --> 02:11.520
The secret is out.

02:11.520 --> 02:13.220
I am, I'm an anesthesiologist.

02:13.320 --> 02:20.300
Uh, I work at Mass General Hospital and I'm also the medical director of biomedical engineering for the Partners Healthcare System.

02:20.920 --> 02:30.400
Uh, in 2004, I started a lab working on medical device interoperability and a few years ago, added cyber security to our portfolio.

02:30.400 --> 02:35.900
And we have a lab which we have no patients of either kind.

02:36.760 --> 02:40.240
Uh, and we, so we have a simulation environment .

02:40.240 --> 02:50.900
We have a sandbox, uh, in which we can look at different scenarios and what the clinical impact might be if either a device or if a part of a system, uh, is affected.

02:50.900 --> 02:57.570
Our work has been supported, uh, mostly through federal grants, um, including work supported by the FDA.

02:57.900 --> 02:59.140
So when Dr.

02:59.140 --> 03:08.000
Schwartz is, uh, passed on, uh, the mic is passed on to her, um, we've been collaborating closely on medical device cyber security and also with other agencies.

03:08.100 --> 03:23.280
Uh, in order to provide information and resources, um, education, knowledge, uh, and work together as a community to address what we know is, uh, a serious threat, uh, and to avoid, um, any negative effect to patients.

03:23.280 --> 03:25.900
And that could happen if a device is affected.

03:25.920 --> 03:28.100
It could happen if a device isn't available.

03:28.100 --> 03:33.300
It also could happen if, if incorrect information is disseminated to clinicians and it overwhelms them.

03:33.300 --> 03:38.780
So, you know, like any complex ecosystem, uh, it's really difficult to address all the issues.

03:38.780 --> 03:45.640
But we're working with a lot of, uh, collaborators in trying to do that and we're really looking forward to the conversation today.

03:46.240 --> 03:47.760
Thank you, Julian.

03:47.760 --> 03:53.240
So I'm Suzanne Schwartz and I'm at FDA's Center for Devices and Radiological Health.

03:53.340 --> 04:03.820
I work, uh, as the director of a new office as a result of a reorganization , the Office of Strategic Partnerships and Technology Innovation at CDRH.

04:03.920 --> 04:21.680
And it is a pretty expansive portfolio that we have within this office, all extremely exciting, leading edge work, uh, which makes every day an extraordinary challenging, but again, uh, intellectually, scientifically stimulating and exciting, uh, portfolio.

04:22.140 --> 04:26.420
Included within that portfolio is medical device cybersecurity.

04:27.720 --> 04:32.060
Interestingly enough, um, my background, I'm a clinician as well.

04:32.060 --> 04:33.100
I'm a physician.

04:33.100 --> 04:37.500
My training, um, has nothing to do with cybersecurity.

04:37.640 --> 04:57.680
I'm a burn surgeon actually, uh, and, uh, prior to coming to the FDA, what will be nine years ago, uh, in October, I worked both within the clinical and academic sphere as well as sometimes spent in the private sector as well, working within the medical device space for a startup technology,

04:57.680 --> 04:58.600
uh, company.

04:58.620 --> 05:16.260
But when I came to the FDA as a medical officer, uh, I was placed within CDRH and, uh, was involved really in the pre-market side of review and consultation on submissions of new technologies that has to do with my field, burn injuries and wound repair,

05:16.260 --> 05:17.620
really exciting.

05:17.860 --> 05:42.100
Um, as a result of, long story short, as a result of the work that I was doing, I was introduced to the field of what's called emergency operations and medical countermeasures at FDA and, uh, was assigned to really represent efforts on behalf of the center for public health related emergencies.

05:42.200 --> 06:04.420
And when we were struck several years ago, this goes back to 2013, with the presentation of vulnerabilities of medical devices to the agency and really were not sure what we were supposed to do with that information, especially when it comes in as a huge package of information,

06:04.420 --> 06:08.780
um, through the researcher and through the Department of Homeland Security.

06:08.780 --> 06:21.520
The agency had made a decision at the leadership level that it's within this public health response role that we would undertake efforts for medical device cyber security.

06:21.520 --> 06:32.620
So what started out, the point here is what started out as purely reactive and response, subsequently evolved and morphed into a much more proactive forward leaning stance.

06:32.620 --> 06:42.960
And it took really putting together a team across the center of experts that really represent the entire product life cycle.

06:42.960 --> 06:47.980
But beyond that, it was really about recognizing what we lacked.

06:48.080 --> 07:03.960
And that's a really important point for our discussion here because it's recognizing that there is a lot of expertise, knowledge, resources that the agency did not have and still does not have with respect to medical device cyber security.

07:03.960 --> 07:06.720
So what do you do in that set of circumstances?

07:06.760 --> 07:08.160
You reach out.

07:08.340 --> 07:10.300
You collaborate.

07:10.300 --> 07:11.360
You look for partners.

07:11.360 --> 07:24.020
You look to understand exactly what it is that we need to be working on together as a community in order to address some of these really complex challenges and this evolving space.

07:24.020 --> 07:30.780
And it's really as a result of that over the past several years that we've come a significant distance.

07:30.780 --> 07:32.460
We still don't make any mistakes.

07:32.460 --> 07:36.460
We have a long, long way to go as an agency and as a community.

07:36.460 --> 07:40.020
But it is an arduous journey.

07:40.400 --> 07:58.240
But it's through the types of collaborations and partnerships that we've had through your community, through the hacker community, with industry, with healthcare delivery organizations, with patients, with others, that has brought us to where we are right now in articulating policy,

07:58.240 --> 08:07.360
guidance, in defining collaborations and understanding what are ongoing challenges and we'll continue to do that as time goes on.

08:07.360 --> 08:08.400
Thank you.

08:09.960 --> 08:11.600
Hi, I'm Dave Guffery.

08:11.600 --> 08:14.300
I'm the biomedical cyber security specialist for Partners Healthcare.

08:14.300 --> 08:19.880
I lead the medical device cyber security program for operations and research across the healthcare enterprise.

08:19.880 --> 08:30.440
Background is kind of winding and varied, so I started out doing electrical stimulation feedback systems and signal processing non-linear and linear systems.

08:30.460 --> 08:37.800
Transitioned into doing virtual reality as well as brain computer interfacing and other kinds of sensory motor integration.

08:38.060 --> 08:56.820
Came to Partners Healthcare a number of years ago, was on the medical device integration team, leading architecture development, medical device integration across the enterprise with our health record system and overseeing cyber security as well as monitoring systems pieces for the medical device integration.

08:57.220 --> 09:03.720
And a few years ago, transitioned to this current role, leading the medical device cyber security program across the enterprise.

09:03.720 --> 09:26.640
So I work organizationally across all of our sites, all the different information systems groups as well as our clinical groups, biomed groups and conduct research in the medical device interoperability cyber security program lab that is co-run with Julian Goldman over here who is our medical director of biomedical engineering and in conjunction with our chief information security officer.

09:26.640 --> 09:34.140
We do a lot of work across the organization for both operations and research within the lab and we'll talk about some of those activities here today.

09:34.480 --> 09:36.840
Give it up, what a panel, huh?

09:37.420 --> 09:39.880
I'll keep it for a minute and pass it around.

09:40.520 --> 09:43.640
Alright, I'm waking up.

09:44.800 --> 09:48.260
I think we're going to talk about a lot of serious problems today, right?

09:48.260 --> 09:54.940
We're going to be talking about some of the big difficulties that still remain in the space, issues of patient safety, etc.

09:55.100 --> 10:03.660
But before we do that, I think it's important for us all to take a little uh history lesson if you will and look back at how far we've actually come.

10:03.660 --> 10:11.820
So being involved in this space for, for several years now, I remember first coming to this and thinking , wow, everything is broken, it's a giant dumpster fire, right?

10:11.820 --> 10:12.820
How many people have heard that?

10:12.820 --> 10:15.680
Like healthcare, cybersecurity, medical devices, a giant dumpster fire.

10:15.780 --> 10:21.460
You know, I don't know how raging it is right now, I mean there's probably some varying opinions, could you still roast some marshmallows?

10:21.460 --> 10:22.120
Probably.

10:22.660 --> 10:27.200
But I think it's important for us to take a little bit of a trip down memory lane.

10:27.200 --> 10:30.760
I'm going to ask each of these panelists to just reflect briefly on how far we've come.

10:31.240 --> 10:33.100
Uh please, if we can start.

10:33.320 --> 10:42.980
And just talk about how far the space has come, what have you seen in the last 10 years or so that gives you some hope about what we've accomplished before we just tear it all down with the negative stuff.

10:44.120 --> 10:44.920
Sure.

10:45.720 --> 11:05.220
I think, you know, regarding how big the fire is and is it only sufficient for roasting marshmallows or, or what, I think put into perspective that when we talk about medical device cybersecurity, we have to kind of think about which part of the system we're talking about.

11:05.220 --> 11:08.260
Are we talking about a device itself and devices being affected?

11:08.260 --> 11:12.760
Are we talking about men in the middle attacks of data that might flow within a health system?

11:12.760 --> 11:16.880
Are we talking about, for example, ransomware of, of health records?

11:16.900 --> 11:22.640
And probably the answer and where we are with that differ depending upon which part of the system that we're talking about.

11:23.740 --> 11:41.560
But I think, you know, just even before digging deeper into that and we'll all do that, I think everyone should have a small sense of comfort knowing that medical devices have always been imperfect and they've always have the potential for failing.

11:41.560 --> 11:43.540
I mean, that's what equipment does.

11:43.580 --> 11:54.060
And so the basic part of, of medical training, of, of clinical training, nursing training, biomedical engineering training is to deal with equipment that isn't performing correctly.

11:54.060 --> 11:56.020
Things break, they don't work.

11:56.160 --> 12:04.640
And so depending upon the criticality of, of a medical procedure, there are backup systems in place and always have been.

12:04.640 --> 12:11.560
And as an anesthesiologist, for example, we've always expected that the anesthesia machine and ventilator will fail during an anesthetic.

12:11.560 --> 12:13.520
We always expect that our other equipment will fail.

12:13.520 --> 12:19.260
We've always thought that way ever since these things were developed and when they used to fail a lot more frequently than they do today.

12:19.540 --> 12:21.480
So the reasons for failure are different.

12:21.480 --> 12:24.720
They used to simply, you know, be mechanical failures, pneumatic failures.

12:24.720 --> 12:30.020
And now that everything is microprocessor based, we have to be more concerned about cyber security.

12:30.020 --> 12:36.160
But at least so people, you know, don't cancel the surgery that's scheduled for tomorrow or next week.

12:36.340 --> 12:39.300
Be aware that, you know, failures happen.

12:39.320 --> 12:47.920
Now, we're not used to seeing large fleet failures of entire, you know, products across the whole hospital, for example, which of course is the big fear with cyber security.

12:47.980 --> 12:53.020
However, those things have happened also completely unrelated to cyber security.

12:53.020 --> 13:04.360
For example, if there's a defect in the plastic tubing that's used for, for infusions of medication and, you know, all that truckload of that defective tubing arrives, it's all distributed throughout the hospital.

13:04.360 --> 13:07.020
The next thing you know, all the pumps are affected.

13:07.300 --> 13:12.100
There have been software upgrades over the years that have caused problems with medical devices, critical devices.

13:12.100 --> 13:14.560
So even that is not entirely new.

13:14.580 --> 13:21.460
But in a sense, the vector and the nefarious nature is what's changed probably the most.

13:21.680 --> 13:25.800
And so I think the end result of this is we're talking about it now.

13:25.800 --> 13:26.860
We didn't ten years ago.

13:26.860 --> 13:28.360
We're thinking about it now.

13:28.600 --> 13:37.600
And what we have yet to do is start to ensure that everyone has a higher level of vigilance for cyber security related device failures.

13:37.680 --> 13:41.180
And we need new ways of being able to address those when they occur.

13:41.900 --> 13:44.940
So instead of going on for an hour, I'm just going to stop right there.

13:47.300 --> 14:06.540
Julian, I'm going to pull on one thread that you mentioned because I think when I reflect on the past several years, I think a lot of it comes down to a significant change in culture a significant change in mindset and how we think about cyber security.

14:06.640 --> 14:22.260
And what I mean by that is an example or illustrative of where we are now versus where we were several years ago is that in many of the conversations that we would have, FDA, whether it's with industry, whether it is with healthcare delivery organizations,

14:22.260 --> 14:28.240
whether it's with just general public, it was, you know, kind of, that can never happen.

14:28.240 --> 14:30.600
That's just purely hypothetical, theoretical.

14:30.600 --> 14:32.920
I don't need to be worrying about that.

14:32.920 --> 14:39.320
And how that would translate with respect to how even industry would view an issue that we'd bring up.

14:39.320 --> 14:47.720
It was like, you know, we don't design our devices to be thinking about the potential mal-intent or nefarious intent.

14:47.720 --> 14:50.340
We're designing for intended use.

14:50.340 --> 15:03.460
So think about the shift in mindset that we've asked of industry in the design of new devices to be putting on the hat of the adversary, to be thinking about it from the standpoint of threat modeling.

15:03.460 --> 15:16.500
It's like, no, we need to consider other potential mechanisms by which a device could be used other than what it was perhaps, you know, in the most beneficent sense intended for.

15:16.500 --> 15:33.140
The same mindset shift was required of healthcare organizations because, and of providers, because all of us who work within this field, it's hard for us to wrap our brains around the idea that anybody would want to do anything bad.

15:33.280 --> 15:38.780
That why would you look to disrupt operations within a hospital?

15:38.780 --> 15:40.940
Most vulnerable of individuals there.

15:40.940 --> 15:42.160
Why would you do such a thing?

15:42.160 --> 15:44.040
Nobody would ever do such a thing.

15:44.040 --> 16:03.400
But the recognition over the years, first of all, that hospitals, you know, in some ways they represent that they are a soft attack surface and they therefore have to be thought of as kind of a hostile environment in terms of attempts at attacks on the systems.

16:03.400 --> 16:10.380
And CISOs of hospitals will tell you that that happens and that's been happening day in, day out, all the time.

16:10.380 --> 16:15.440
That they are constantly trying to defend from attacks, attempts at intrusions.

16:15.440 --> 16:40.240
So when you think about the requirement of taking on a very different posture here, that means really moving to a different place with regard to what, first of all, FDA would expect of manufacturers in the design of devices as well as in the ongoing vigilance and management and monitoring of these devices throughout their lifetime.

16:40.240 --> 16:50.760
And what the expectations are, not only in the monitoring, but then therefore in attending to vulnerabilities that do get identified through the lifetime of that device.

16:50.760 --> 17:08.460
And we cannot just ignore those or leave those, but they really do need to be looked at, assessed, validated, assessed, and determined in terms of what the potential is for patient harm for a safety concern coming out of that vulnerability were to be exploited.

17:08.600 --> 17:19.060
So those are some prime examples of what, at the very core, the very foundation is required in order to move the ecosystem.

17:19.060 --> 17:23.960
There needs to be an acceptance of changing that culture, changing that mindset.

17:23.960 --> 17:25.520
It is an ongoing process.

17:25.520 --> 17:42.260
I think that we have certainly seen a lot of understanding of this and buy-in from industry as well as from the hospital health care side, particularly, again, from CISOs that are on that front line, where there's a lot of work that's yet to be done.

17:42.260 --> 17:58.100
And I think that this will be a good, also, segue as we talk about today's discussion in medical simulation where there's a lot of work that has to be done is with the clinician community, clinicians and patients as well, but clinicians who are, again,

17:58.100 --> 18:07.260
focused on treating patients and giving care and thinking about things in the most kind of beneficent manner.

18:07.260 --> 18:28.600
The notion that something can, something, someone can impact on a device's safe performance and, therefore, considering the benefit-risk calculus, understanding what that is, and how to have dialogue with patients and how to also be on the lookout for that,

18:28.600 --> 18:30.660
that is a whole new world.

18:30.780 --> 18:39.140
And that's a world that, you know, Christian and I have, we've talked a lot about, and that is where we need to be thinking about as we go forward.

18:39.360 --> 18:55.200
Medical simulation offers a real, you know, extraordinarily, really excellent tool in terms of being able to do that, particularly because it's something that resonates with the clinician community, and we'll talk about that as we go forward.

18:56.720 --> 19:04.600
I would say where in my tenure, my tenure hasn't been as long in the field as the other two well-respected people next to me.

19:04.640 --> 19:11.420
However, what I have seen in this time frame has been a rapid increase in connectivity.

19:11.420 --> 19:15.140
So, medical devices now are not functioning as an island.

19:15.140 --> 19:18.960
We don't even have vendor systems individually functioning as an island.

19:18.960 --> 19:34.980
You have a system of systems that are connected together with, you have imaging, you have clinical monitoring, you have electronic health record systems, you have laboratory systems, we have even R&D systems, they're all tied together in some manner, all on our networks.

19:34.980 --> 19:57.300
And because of this rapid connectivity increase over the last number of years, as a result of several governmental initiatives as well as other aspects that have gone into this, this threat surface is, is increasing with that and the awareness of cyber security and the need for cyber security is increasing in the medical community.

19:57.300 --> 20:10.320
We have individuals such as myself now popping up at other health institutions across the country where you're having medical device specific experts devoted to securing their systems within the healthcare organizations.

20:10.360 --> 20:16.720
We have procurement aspects that are going into, you're considering cyber security now as part of your procurement process.

20:16.720 --> 20:25.060
You're assessing your medical infrastructure intake into the company prior to deployment and you're making recommendations and mitigations as a result of that.

20:25.060 --> 20:44.540
You're reassessing your existing inventory that may have been, your medical device systems may have been in operation for decades potentially and you're looking at it with a new lens to see where are your potential weaknesses, how can you fill them and where do we go for the future so that we can make better decisions?

20:46.240 --> 20:46.880
Awesome.

20:46.880 --> 20:49.840
So of course we picked the best of the best, right?

20:49.840 --> 20:58.320
The people most aware of this uh, representing institutions that are very savvy when it comes to medical device critical hospital infrastructure security, right?

20:58.320 --> 21:00.140
So the best of the best, right?

21:00.300 --> 21:03.140
I'm gonna ask a provocative question to the panelists here.

21:03.140 --> 21:11.180
Put, put on the hat of a uh, middle tier, somewhat rural hospital um, that takes care of 100,000 patients a year.

21:11.180 --> 21:18.060
So still takes care of a lot of patients but isn't connected to a big hospital network, doesn't have tremendous amounts of resources, etc.

21:18.500 --> 21:34.320
And take what you know about how hard it is to secure these systems, how vulnerable it perhaps could be and answer the question, do you think that if a patient was harmed by a device that was compromised by a piece of malware or some other thing, that we would,

21:34.320 --> 21:37.520
that hospital would be sophisticated enough to detect it?

21:39.530 --> 21:42.870
I think that's a very tough question to answer.

21:42.970 --> 21:59.610
Where I would start with it is, it depends on A, the monitoring infrastructure that they have, B, the relationships that they have with the medical device manufacturers themselves and C, the technical personnel that they have on staff.

21:59.930 --> 22:05.010
A, this is a mid-sized hospital, not a small rural community hospital.

22:05.010 --> 22:21.370
So you can propose that they are going to have some technical resources, you are going to have some form of network monitoring, you are going to have clinicians that are going to be able to see changes in behavior of your medical infrastructure.

22:22.330 --> 22:37.370
And we have many manufacturers in the field that are on top of their game and they are, they will actively work directly with the hospital to help remediate and mitigate their risk.

22:37.450 --> 22:43.870
So as long as those relationships exist, I would think that you can't, it's going to minimize the potential impact.

22:43.870 --> 22:47.490
I can't, it's, I can't directly answer it fully.

22:47.650 --> 22:50.970
No, it's a hypothetical, it's my favorite one because you can't answer them.

22:52.250 --> 23:08.830
I would be pessimistic about the ability of such a hospital being able to recognize a cyber security related incident as being just that.

23:09.070 --> 23:15.570
And I think it has a lot to do with the fact that the hospital is not well resourced.

23:15.570 --> 23:45.530
It is in perhaps an environment or a location that is geographically well removed from some of the bigger academic or bigger clinical centers where there is a lot more likely exchange and dialogue happening across the C-suites and the operational folks within those organizations.

23:45.530 --> 24:05.370
And we've heard that, you know, there's been an analysis really done of that several years ago through the health care cyber security task force actually that kind of surveyed what does the nation look like with respect to its current state of preparedness and capability.

24:05.370 --> 24:24.910
And there is a very wide spectrum across the United States for hospitals and health care facilities which leaves us certainly very concerned about those types of facilities and how we as a collective, as a community, can not just government but really everyone,

24:24.910 --> 24:46.790
the private sector as well, think about the notion of partnerships and other efforts creatively that can help those types of institutions that may be, to assist when these types of incidents do occur.

24:47.290 --> 24:48.810
Thank you.

24:49.370 --> 24:57.450
Well, I think we should consider what kind of issue we're dealing with, what kind of attack it is.

24:57.450 --> 25:02.490
Is it something that's intentionally in stealth mode or is it something that's intentionally not?

25:02.490 --> 25:18.910
So if the Windows desktop screens pop up a ransomware notice with a very long hard-to-remember Bitcoin, you know, information, I think, sure, the hospital's going to know that's been attacked.

25:19.110 --> 25:35.430
If on the other hand someone is performing a man-in-the-middle attack to modify lab values and other data with an intent to assassinate someone or cause harm or disrupt operations, I imagine it would be very difficult for anyone to detect.

25:35.690 --> 25:50.090
If a medical device is bricked, if the medical device obviously isn't functioning correctly, what people will do is grab another medical device, call their biomed department, clinical engineering department, call the manufacturer and say, hey, this thing's a piece of junk,

25:50.090 --> 25:50.970
please fix it.

25:51.090 --> 25:56.310
Manufacturer might examine it and hopefully would identify that it was a malware-related issue.

25:56.390 --> 26:03.910
If it happens to many devices, as soon as they're connected to the network, I think the light bulb will come on pretty quickly that something isn't right.

26:04.090 --> 26:10.670
And probably most hospitals would, you know, one of the things they would do, of course, call the manufacturer and hopefully the manufacturer would know.

26:10.670 --> 26:13.930
And a lot of small medical device manufacturers aren't resourced.

26:14.270 --> 26:18.430
You know, probably to sort this out easily , there are a lot of mom and pop operations.

26:18.450 --> 26:25.010
Others that are larger are well resourced, but they'll often call up the FDA and they'll call Dr.

26:25.010 --> 26:27.990
Schwartz and say, we got a problem, could you help us out?

26:27.990 --> 26:28.530
What do you know?

26:28.530 --> 26:30.290
What does the intel look like?

26:30.290 --> 26:37.550
So I really think it could cover the whole spectrum and to some extent depends upon how stealthy they're intending to be.

26:38.530 --> 26:41.090
I'm not looking forward to that day.

26:42.750 --> 27:00.250
There's a lot of talk about how we need to do a lot of technical work and process improvement from when a security researcher finds a vulnerability in a device to when they interact with the medical device manufacturer, interact with the regulators, to perhaps when they put in mitigating controls,

27:00.250 --> 27:01.130
perhaps patches.

27:01.130 --> 27:05.470
I mean, we've spent a huge amount of the last 10 years talking about that process.

27:05.890 --> 27:10.790
We're going to transition a little bit into talking about this unknown part.

27:11.130 --> 27:17.450
Just to put a little bit of background out there, imagine an implantable medical device needs a patch.

27:17.930 --> 27:20.870
This patch doesn't go over the air like your phone.

27:21.030 --> 27:23.370
It's not a very easy thing to deploy.

27:23.370 --> 27:27.390
In fact, it almost always has to be deployed by a doctor in a doctor's office.

27:27.570 --> 27:28.810
Why, you ask?

27:28.830 --> 27:36.250
If it should fail, then the device won't work and you'll need a doctor or nurse nearby to make sure you survive without the function of the device.

27:36.250 --> 27:37.390
Does that make sense?

27:37.790 --> 27:46.130
Many people out there who are probably new to this space in the medical device patching world can already understand how hard it must be to patch these things.

27:46.130 --> 27:51.290
If every time a device has to get patched, you have to go see your doctor, it's a big deal, right?

27:52.350 --> 28:12.750
Now the question to the panelists are, how equipped do you think doctors and nurses, nurse practitioners, the clinical staff is today at A, understanding this, and B, being willing to deploy something like a patch, knowing nothing about cyber security or patient care implications.

28:12.750 --> 28:16.450
So the question again, how well do you think doctors are prepared now?

28:16.790 --> 28:19.630
And then we'll get to in the future how we can maybe change that.

28:20.690 --> 28:23.150
That's a good question, an interesting one.

28:23.430 --> 28:25.030
These things are happening already.

28:25.030 --> 28:31.430
We're receiving notifications that devices might be affected and we contact specialty groups.

28:31.430 --> 28:46.130
For example, if a device, you mentioned specifically an implantable device, but it could be something that's connected and not physically implanted, like an insulin pump, which is similar in some regard, although it's not life-sustaining, the pump itself isn't,

28:46.130 --> 28:48.310
because someone could use an injection instead.

28:48.330 --> 28:55.350
When we contact our different physician groups and others, there's really no pushback.

28:55.430 --> 28:59.610
They think they're used to being advised by other experts when it comes to technology.

28:59.730 --> 29:01.670
And they ask, what is the risk?

29:02.490 --> 29:07.650
They rush to find out if the devices they have are affected or if their patients are affected.

29:07.930 --> 29:10.130
And then they ask for guidance on what to do.

29:10.270 --> 29:25.050
Now, I don't think we've hit times yet where it looks like something has a very high risk, if not addressed, and a very high risk to the patient if addressed, for the reasons you mentioned.

29:25.050 --> 29:28.950
You know, you don't want to destabilize something.

29:28.950 --> 29:45.450
If you have a device that's working, and let's say is a very low risk that it will be affected, as long as it's not connected to something, if that's the case, then the best course could be transparency, advise everyone, and then make sure that you monitor the situation.

29:45.490 --> 29:52.070
And based on the experience to date, I'm finding very receptive groups of clinicians.

29:52.070 --> 29:58.850
I'm an anesthesiologist, as I said, but I work in my role in biomedical engineering, work with all the other groups across the hospital.

29:58.850 --> 30:07.560
And there's really been no pushback, but sometimes the guidance is not necessarily very clear when we're in the stage of uncertainty, which is usually that first stage.

30:09.230 --> 30:12.330
So I have a slightly different perspective.

30:12.690 --> 30:39.330
I'm having to deal with the responsibility from the FDA of putting forward recommendations through communication to providers, to clinicians, and dealing with what one could consider a much more high stakes or high consequence type of situation where,

30:39.330 --> 30:56.550
as an example that Christian provided, it would be necessary for the physician to bring a patient back in and have a discussion with the patient over what it is that's about to happen and why.

30:56.550 --> 31:01.350
You know, the physician feels that it is a prudent step to take.

31:01.730 --> 31:15.570
And we have seen a lot of reluctance, a lot of hesitancy within certain clinical groups, within certain clinician communities.

31:15.710 --> 31:27.510
The one that is most prominent, and I think it's because also it's the first one that we've really had to deal with, has been within the cardiac space, specifically with electrophysiologists.

31:27.910 --> 31:53.630
And I think that it's a demonstration of really where we are in, you know, the nascency and what will be growing pains here as far as better arming clinicians with what we can anticipate in the future with respect to these types of devices definitely needing to be updated over their lifetime.

31:53.690 --> 32:12.590
And taking that from the standpoint of, oh my god, this is going to create a lot of anxiety and concern and fear for the patient as well as for the physician to something that becomes more of the norm.

32:12.590 --> 32:35.090
That a physician can expect that over the lifetime of a device, the device just like a phone or any other technology, will need to receive updates in order to address the vulnerabilities that get identified later on that are not known or have not been discovered earlier on.

32:35.290 --> 32:55.890
That that becomes, and again this comes back to the shift in mindset and culture, like that this becomes, yeah, part of what the practitioner can expect and therefore how do we ease that level of comfort for the practitioner so that they also know how to have a discussion around that with a patient.

32:58.030 --> 33:32.990
So I think that there's a lot of opportunity for us to talk about approaches that we can take to better educate clinicians, to better inform, to arm them for these situations that we would expect are going to become more normal and that we're not dealing with a kind of crisis every time a vulnerability is discovered within a device that is going to necessitate the physician to take a specific kind of action.

33:33.290 --> 33:36.190
And I'll take it just a step back as well.

33:36.590 --> 33:57.990
I don't think that at this stage for different clinician groups, there's sufficient enough even understanding of cyber security to recognize with whom, with which patients they may feel the action or recommendation is appropriate and where they may feel,

33:57.990 --> 34:10.830
nope, it's not necessary for this particular patient and how to have that kind of benefit-risk related decision, benefit-risk, excuse me, discussion in order to come to an informed decision.

34:10.830 --> 34:25.530
One has to have ample background and understanding of what the risk is and what the consequences are in order to be able to render that kind of a decision together with a patient.

34:25.530 --> 34:27.810
But we're not even at that stage yet.

34:27.810 --> 34:43.070
So simply handing over a set of recommendations or next steps to a physician without having more contextual understanding I think is right now a gap area that needs to be addressed.

34:45.070 --> 34:49.150
I'm in full agreement with the other two panelists.

34:49.150 --> 34:55.630
I think that there is definite room for improvement regarding the educational process and communication process.

34:55.630 --> 35:04.950
So over time there is going to need to be a shift that medical devices will need to be patched just part of normal course.

35:05.250 --> 35:07.870
Eventually we're going to find vulnerabilities in almost everything.

35:07.870 --> 35:09.130
That's just how it's going to happen.

35:09.130 --> 35:10.730
Everybody in this room knows that.

35:10.870 --> 35:15.870
So those kind of conversations need to be starting to happen with the clinical staff.

35:16.230 --> 35:26.490
Now that kind of piece can go into well how do you develop an educational structure within your organization as well as how do we develop a national educational structure.

35:26.670 --> 35:48.930
And those two pieces all three of us on the panel are actively working on and there's a number of people in the room that are working on how do you put out information in a digestible manner without increasing the anxiety while improving the context and awareness of your device and its potential impacts and situations.

35:49.890 --> 35:50.650
Awesome.

35:50.650 --> 35:54.170
So we're going to go a little bit faster through some of the next topics.

35:54.410 --> 36:03.090
Who here in the room would like to fly in a plane with a pilot who has never actually flown a real plane and let them take off and land?

36:03.090 --> 36:04.430
Please raise your hand.

36:05.010 --> 36:06.550
Where's your sense of excitement?

36:06.710 --> 36:08.310
How much to take it?

36:10.550 --> 36:12.150
We could discuss that.

36:12.470 --> 36:14.750
Let's see how much you would pay.

36:14.750 --> 36:15.370
Are we willing?

36:15.370 --> 36:17.330
You know maybe $50 is not enough.

36:17.330 --> 36:20.630
Anyways, we're going to talk about simulation next.

36:20.630 --> 36:23.350
So we're going to have two different types of simulation.

36:23.350 --> 36:26.250
One is simulating attacks on medical devices.

36:26.250 --> 36:29.650
Probably best exemplified for their work at the partner's lab.

36:29.650 --> 36:32.550
So how do we simulate these types of attacks?

36:32.550 --> 36:39.390
How do we put clinical risk and the criticality of these types of things and see how they translate to impacting patients?

36:39.390 --> 36:41.230
That's one type of simulation we're going to talk about.

36:41.370 --> 36:44.270
There's also this other simulation which falls into the pilot part of it.

36:44.270 --> 36:53.230
So, you know, aviation figured this out decades ago that they didn't want pilots flying in planes the first time while learning to fly planes so they put them in these high fidelity simulators, right?

36:53.230 --> 36:57.930
Like basically video games to teach them how to take off, land, push all the buttons the right way, etc.

36:58.290 --> 37:04.610
We stole that in medicine and we teach doctors how to take care of really sick patients using simulation.

37:04.610 --> 37:09.490
Who here wants a doctor who looks like they're 25 taking care of them when they have their first heart attack?

37:09.490 --> 37:10.110
Anyone?

37:10.230 --> 37:12.410
Never having taken care of a heart attack before?

37:12.470 --> 37:13.290
You probably don't, right?

37:13.290 --> 37:25.530
So we got to teach those doctors how to take care of those sick patients using mannequins and dummies and fake emergency departments so that when they give the wrong drug or give too much of a drug or too little, we can correct them before they do it on a real patient.

37:25.530 --> 37:26.910
That sounds great, right?

37:27.750 --> 37:31.530
We did this for, uh, for medical devices, right?

37:31.530 --> 37:43.270
So we took some real research out of the security researcher space and we wrote clinical simulations based on if this pacemaker vulnerability is real and it impacted the patients, this is how it would look.

37:43.530 --> 37:58.650
The patient would come in complaining that their pacemaker kept shocking them and perhaps they'd even go into a deadly heart rhythm as a result or maybe if the infusion pump they were hooked up to was compromised it would give too much or too little medication and we let these doctors go through these cases not knowing what was going to happen.

37:58.650 --> 38:01.930
It's kind of like Doctor Dungeons and Dragons, right?

38:01.930 --> 38:03.790
They unroll the scenario.

38:04.510 --> 38:07.190
I bring all that up to say, Dr.

38:07.190 --> 38:12.250
Schwartz, you saw some of these simulations in a couple of these conferences in the past and what were some of the insights you saw?

38:12.250 --> 38:13.170
Did you see anything?

38:13.170 --> 38:13.970
Was it magical?

38:13.970 --> 38:14.790
Was it boring?

38:14.810 --> 38:24.670
What were some of the real importance to this that you saw that we could leverage to address the very issues this panel is talking about, which are generally clinicians and clinical staff don't know much about this and need to be better informed?

38:24.710 --> 38:48.610
Yeah, I have had now the opportunity, really the privilege, to attend and to observe two such types of summits that went through these types of simulations and for each of the scenarios, observing the clinicians kind of go through step-by-step their interventions,

38:48.610 --> 39:08.510
their kind of watching them think through and act at the same time and then do a debrief afterwards and listen to their aha moments was just extraordinarily insightful and impactful because each of them had said, you know, again, going into such a simulation,

39:08.510 --> 39:33.190
not knowing what the scenario was going to be, that the consideration of an exploit of an attack never would have entered their mind and it brings the understanding that this can occur into a frame of reference that resonates for the clinician, you know,

39:33.190 --> 39:45.630
in terms of what it is that they need to be thinking about in dealing with a emergency or scenario that's unfolding right before their eyes.

39:46.030 --> 40:19.130
And I think it was in some ways an awakening and an enlightenment for each of the clinicians that were involved in those scenarios and for me very much validating that in order to bring better awareness to cybersecurity for clinicians, placing a clinician in an immersive type of learning environment in the way that we do teach and train medical students and house staff and so forth is a very,

40:19.130 --> 40:21.910
very effective method of doing so.

40:22.010 --> 40:39.750
Because the intent is not, again, to turn clinicians into cybersecurity subject matter experts, but rather to give them the tools to be recognizing and to therefore be doing what they do best as a situation is presented to them.

40:40.050 --> 40:47.270
And it was, again, a very, very enlightening and informative exercise.

40:49.630 --> 40:51.810
So if, may I add?

40:51.810 --> 41:03.290
So what is called high fidelity simulation is what this is usually called in the medical education world and a number of specialties now use these environments.

41:03.290 --> 41:10.010
I go through simulation training every few years as well and, you know, boy it makes you sweat.

41:10.190 --> 41:13.050
And, you know, so what's special about simulation?

41:13.050 --> 41:26.270
Well, you can simulate things that someone may never see clinically and you can control and contrive whatever you want so that it becomes not only that educational opportunity, but it also becomes seared in your memory.

41:26.470 --> 41:40.990
And then when there is an event, there's a problem, people usually fall back on, you know, their most memorable events of, oh yes, there was a simulation on this and maybe this issue is related to cybersecurity, you know, would be the outcome of that.

41:41.030 --> 41:44.670
Also a big part of simulation is the team training aspect.

41:44.850 --> 41:55.710
It's learning what to do when devices fail, things fall apart, and it's learning how to recognize that something is actually much worse than it first appeared to be.

41:55.710 --> 41:57.910
Because we've all had bad things happen.

41:58.210 --> 42:04.610
Could it be a fire in the kitchen or it could be a car accident and at some point things look okay and then they suddenly become terrible.

42:04.790 --> 42:10.410
And we train clinicians to recognize that tipping point, which is, you know, hard to do.

42:10.410 --> 42:13.870
You have to kind of step out of the experience and know how to bring in other help.

42:13.870 --> 42:25.630
So there's a lot of value at multiple levels for simulation and I think it's been really great to, you know, go down that pathway because of the potential for all sorts of good things to come out of it.

42:29.250 --> 42:42.510
And you're of course welcome, Dave, to comment on that or if you'd like to talk more about your simulating of hacking these devices because the last topic I'd like to hit after you talk about that is really how do we involve this community in here because they're hackers out there,

42:42.510 --> 42:42.770
right?

42:42.770 --> 42:44.970
We care about being involved in this.

42:44.970 --> 42:46.070
That's why they're out here.

42:46.070 --> 42:54.930
They are patients themselves or their parents or their kids are patients and they want to help but they don't fall under the traditional categories, right?

42:54.930 --> 42:56.790
They're not in academia.

42:56.970 --> 42:59.150
They don't care about writing papers.

42:59.150 --> 43:00.330
They want to hack.

43:00.330 --> 43:04.890
And there are a lot of barriers to hackers out there getting involved in this space.

43:04.890 --> 43:14.210
So please talk about your work in the lab and then also let's start talking about how we can broaden inclusion into this space which is really hard.

43:14.210 --> 43:16.450
Hackers can't get ahold of devices easily.

43:16.710 --> 43:18.210
They don't know who to talk to.

43:18.230 --> 43:23.450
There's a lot of problems and pushback and even threats from certain organizations being involved in that.

43:23.450 --> 43:25.550
So if we can move on to that for the last topic.

43:25.630 --> 43:37.850
Alright, well I'll say one comment on the previous topic of one area in these various simulations that has been very enlightening is learning who needs to be at the table.

43:37.870 --> 43:49.110
So when you're doing emergency preparedness response activities and you're working through emergency situations, it's important to figure out what your communication paths are.

43:49.110 --> 43:51.270
Who needs to be in what particular role?

43:51.270 --> 43:52.890
How do you all work together?

43:52.890 --> 43:56.010
You need to have people from the federal regulation level.

43:56.010 --> 44:01.210
You need to have people from your various groups within your hospital organization.

44:01.210 --> 44:03.370
You need to have medical device manufacturer groups.

44:05.010 --> 44:15.630
You have, so all of these various pieces have come out in some of our simulation activities working through these kind of emergency disaster situations within our laboratory.

44:15.630 --> 44:23.810
As far as other kinds of more hacking oriented research and activities that we've done in the lab.

44:23.810 --> 44:30.830
So we use the medical device cybersecurity lab for a number of different pieces both on the operation side and the research side.

44:30.850 --> 44:38.210
Operationally, we do penetration testing of our systems prior to deployment, prior to procurement.

44:38.250 --> 44:51.710
You work through your various systems and figure out how you can work then with your organization and with your medical device manufacturers to fill any potential gaps and figure out how can you have mitigated controls.

44:52.250 --> 44:57.090
As far, we also have work under a project with the department of homeland security.

44:57.150 --> 45:03.970
And this is an aspect where all of you in the room, I highly encourage you to join us in this effort.

45:03.970 --> 45:07.290
So this is under the DHS impact program.

45:07.290 --> 45:11.210
And if you google impact, it'll pop up.

45:11.210 --> 45:16.030
And we have a project for generation of medical device network traffic.

45:16.030 --> 45:23.450
And the reason for this is because as you have all experienced in this research community, you can't get a hold of easily of medical devices.

45:23.450 --> 45:26.690
You may not be clinical experts and know the clinical workflows.

45:26.690 --> 45:31.610
And there may be other legal barriers to be able to do research on these medical devices.

45:31.810 --> 45:41.090
And because of that, there's a lack in our field of networking systems and understanding of the medical network traffic in order to develop tools.

45:41.090 --> 45:44.850
We have tools in our regular IT networks.

45:44.850 --> 45:45.830
We have visibility.

45:45.830 --> 45:47.290
We have detection systems.

45:47.290 --> 45:48.950
We have artificial intelligence.

45:48.950 --> 45:52.090
All these different pieces on our IT networks.

45:52.090 --> 45:57.470
We don't necessarily have access to that on our medical networks due to a number of different hurdles.

45:57.990 --> 46:07.750
We are doing under this grant, generating medical network traffic under normal state scenarios, failure mode scenarios, attack scenarios.

46:07.750 --> 46:17.950
We're bringing this to the research community so that you can then get access to this traffic and be able to start developing tools to help improve the ecosystem as a whole.

46:19.010 --> 46:20.430
Yeah, that deserves a clap.

46:20.430 --> 46:22.130
Yeah, get the data.

46:22.130 --> 46:23.270
That's awesome.

46:23.310 --> 46:24.570
Build the tools.

46:25.610 --> 46:36.130
I just want to clarify, when Dave said we don't have those tools on medical networks, the we in this case he meant was the healthcare sector.

46:36.130 --> 46:37.850
Not we in our hospitals.

46:38.890 --> 46:56.170
So the very nature of how medical devices are connected in hospitals is they're typically on separate and proprietary networks for a host of things that relate to the fact that they were not necessarily designed to plug into your backbone.

46:56.410 --> 47:03.190
I won't go into the technical details, but the bottom line is that they're typically isolated and separate.

47:03.190 --> 47:06.850
And that's why this is a data set that's been hard to get access to.

47:06.850 --> 47:14.790
Another reason is that medical devices are expensive and one needs a license to buy them, unless you buy them on eBay, in which case you just need money.

47:15.730 --> 47:21.050
And then as Dave said, setting them up requires a certain level of clinical engineering knowledge.

47:21.190 --> 47:38.670
Part of the reason we set up the lab the way we did, that's allowing us to share the data through the DHS impact program, part of the way we set it up was to answer a question of what does a laboratory sandbox need to be prepared for a cyber security event,

47:38.970 --> 47:40.550
a significant one.

47:40.750 --> 47:43.970
Hopefully not cyber Armageddon or whatever it's called.

47:44.350 --> 47:51.010
But these things could happen and we as a organization and as a community have to be prepared.

47:51.050 --> 48:12.590
So part of that work over the last year and a half or so has been a project sponsored by the FDA in partnership with MITRE to build a medical device cyber security sandbox to understand what do we need, what kind of tools, how do we simulate different hospital environments and different network architectures,

48:12.590 --> 48:15.810
different implementations and configurations of equipment.

48:15.910 --> 48:27.470
Manufacturers may dictate that for a device to work correctly or a hospital may have to dictate a configuration requirement based upon their available technology or clinical workflow, physical layout.

48:27.610 --> 48:32.550
They may have antiquated equipment, networking equipment, they may have state of the art equipment.

48:32.550 --> 48:42.090
And so we've been studying that for some time and built out the cyber security sandbox as part of this cyber security preparedness initiative.

48:42.090 --> 49:07.250
And you can see how there's really good synergy between these initiatives, whether it's DHS science and technology, S&T program wanting to share more core data with the cyber security community, whether it's the FDA and Suzanne's office thinking about how do we prepare as a healthcare sector so that the community could rapidly engage,

49:07.250 --> 49:12.270
hopefully ahead of time, not in response, but proactively to these events.

49:15.100 --> 49:16.440
Do you have anything, Suzanne?

49:17.100 --> 49:36.020
I would just simply add that I think that both the types of scenarios that we were talking about, whether it is the clinical simulation for providers or whether we're talking about, again, the test bed or sandbox, they both demonstrate the power of simulation for preparedness.

49:36.020 --> 49:47.260
And we always talk about the importance of preparing, exercising, exercising, exercising, exercise to failure, not to futility.

49:47.260 --> 50:04.660
It's the idea of before something bad happens, let's play this out and understand exactly what our weaknesses are, what are the things that we need to shore up, and then reiterate on that and continue to learn through that process.

50:04.740 --> 50:07.560
Simulation is a really powerful tool to do that.

50:08.820 --> 50:09.440
Phenomenal.

50:09.440 --> 50:11.260
All right, so we're approaching the end here.

50:11.260 --> 50:16.920
I just wanted to, again, thank our esteemed panelists and leave perhaps a challenge to the audience and everyone here.

50:16.920 --> 50:19.380
We've talked a lot about simulation as a powerful tool.

50:19.400 --> 50:28.620
Let's practice, let's tabletop these things, both to educate the doctors and nurses taking care of patients, to talk about involving hackers into this space.

50:28.620 --> 50:33.000
But really I feel like simulation is a tool that can be applied across this community, right?

50:33.000 --> 50:36.760
And that's really something I'd love to see go in a different direction.

50:37.240 --> 50:53.260
We'll continue to do clinical simulations, we'll continue to do the technical simulations, but what we really need is involving more of the community outside the walls of a hospital, you know, tapping into the tremendous resources of all the hackers here and allowing them to experience what this would actually be like.

50:53.260 --> 51:01.480
Because what we want to do is leverage the expertise out in the room, all those hackers, to do meaningful work to help secure patients.

51:01.620 --> 51:07.180
And we talked about all the difficulties of how to do that and all the intricacies and how much you have to know to do that.

51:07.180 --> 51:11.920
Let's reduce those barriers, if you will, to allow more hackers into the community.

51:11.920 --> 51:25.480
And I guess I want to say thank you as well to everyone that's involved in the We Heart Hackers initiative, the biohacking village next door, please sit down and take a stab at the CTF, check out the vulnerable devices, and also join the community.

51:25.480 --> 51:27.900
This is something that's really important.

51:29.300 --> 51:35.080
Medical devices take care of our brothers, our sisters, our moms, our friends.

51:35.400 --> 51:55.740
There are a lot of nuances in this community we should be banding together to solve a lot of these challenges and avoid the pitfalls that have happened in other industries that result in either outcomes we don't like, delays to fixing this issue, and then also it's a cool space to be in.

51:55.800 --> 51:57.500
Who's having fun here at DEF CON?

51:57.500 --> 52:00.680
You came to the simulation panel, you didn't even know what it was.

52:00.740 --> 52:01.640
We had some fun.

52:01.640 --> 52:03.420
Can we give it up for our panelists please?

52:03.800 --> 52:03.880
Thank you.

52:03.880 --> 52:03.900
Thank you.

52:03.900 --> 52:04.320
Thank you.
