0	5680	Thank you all for being here. I'm Andrea Matwishan. I'm a professor at Penn State, and it is my
5680	16020	distinct pleasure to share the next hour with Dr. Suzanne Schwartz of the FDA. And I realize
16020	22840	I owe Anthony Ferrante a mug, because this is a... I promised everyone coffee time. There's
23140	29740	a de minimis mug that shall be yours. Maybe when Q&A starts, I'll produce the mug.
29740	30080	Okay.
30080	38440	Like habeas corpus only. Habeas mugus. I don't know. Whatever. Anyway, so today we'll discuss
38440	48160	with Suzanne some of the current FDA approaches to security and learn a little bit about where
48160	55000	we see things going with respect to medical devices and security. But first, I'd love
55000	60340	to start with just asking Suzanne, how did you become interested in security? You weren't
61080	66960	born doing your role. You arrived in your role. So tell us a little bit about your history
66960	76460	and how you went from prior iterations of self to your current job and your deep engagement
76460	83280	with this area. And I will say, without hesitation, the FDA is probably the most active agency
83280	90340	right now on security. And kudos to the agency on that. And you're looking at one of the
90340	96320	main reasons for the FDA's deep interest and forward-looking approach in security. So I'd
96320	100740	love to hear the story of how you became yourself.
101080	105900	Thank you. And I have to say that, no, I wasn't born into security. I think when I was born,
105900	113200	computer security actually didn't exist. I'm dating myself here. In any case, I'm a
113200	120380	physician and my background and training actually, clinically, is in the area of surgery, general
120380	127990	surgery, specifically burn injuries and trauma, which is far afield from the area of cybersecurity.
128280	134080	I've done a lot of work as both an academic, what's called a surgeon-scientist. I spent
134080	140020	quite a bit of time actually within an academic clinical center in New York City, Weill Cornell
140020	146960	Medical Center, which is part of kind of the New York Presbyterian Complex, doing both burn-related
147760	154440	and general wound repair-related clinical care and research. And one thing I would say just about
154440	166580	myself is the desire to pursue research and innovation and looking at new approaches,
166580	174380	new technologies that can further benefit patients. And that's a common thread really
174380	179360	throughout my entire career. Aside from the work that I had done on the clinical side
179970	186900	in the hospital setting, I've also been on the industry side. Before coming to the FDA,
186900	193380	I worked in the private sector as a medical director of a startup company, a startup that
193380	202140	was involved in actually producing and testing a tissue-engineered skin, skin for burn injuries and
202140	209320	wound repair. And it was just kind of over the next few years, this kind of realization that for
209320	216360	me, being able to straddle multiple sectors of the universe, the research and science side and
216360	223640	technology and understanding of the clinical needs, a desire to see technologies get to the bedside.
223640	229320	And in that role as a medical director, I did a lot of interfacing with the FDA in order to get
229320	235700	clinical trials through and to be part of putting applications together. So there's a certain
235700	243120	nexus there that for me was very appealing in terms of pursuit of what can I do to be more
243120	252700	impactful? What can I do from a more general, broader public health perspective in seeing patients
253420	263560	get new therapies? And so fast forward a bit in 2010, I came to the FDA in a role that's called
263560	267220	the Commissioner's Fellowship. It's actually unfortunately been phased out or it's in the
267220	274580	process of being rebirthed as something a little bit different. But the idea of a FDA Commissioner's
274580	281200	Fellow, we came in as a cohort, were to really bring in individuals with varying subject matter
281200	288440	expertise in different disciplines that may be lacking at the agency. And by bringing in
288440	293000	individuals with different subject matter expertise, certainly that benefits the agency
293000	300520	by helping us with some of these newer, more advanced methodologies and technologies to help
300520	305860	in review of those and an assessment of those. And at the same time, the fellowship would enable
307620	315280	the individual to be exposed to the regulatory framework and what's happening with the FDA. So,
315280	321540	whether that individual then ends up going back out to the private sector and acting or serving
321540	327040	as an ambassador to the environment in which they're going to work, where they understand
327040	335620	how regulation in the medical product area works, that's one benefit. The other benefit is for those
335620	343060	individuals where it was determined to be a really mutually beneficial match. So, for myself,
343060	350400	I was very excited about the work that I was assigned at the FDA when I came as a Commissioner's
350400	360700	Fellow. And I saw a huge opportunity to have impact, very, very challenging and novel emerging
360700	366960	areas and an opportunity to learn more and be that continuous constant learner, which is something
366960	373060	also that's important to me in whatever I do. So that at the end of every day, I feel, wow,
373060	379060	I've learned something new or been challenged with something different and tomorrow's a new day and
379060	385080	who knows what that's going to bring. In my journey at the FDA, although I came in as a
385080	388500	Commissioner's Fellow and then because I'm a physician, that translates to being a medical
388500	398040	officer, because I was also representing a field that was a very unique field that you don't have
398040	404840	that many surgeons around that are burn surgeons. It is a small field, but there were connections to
404840	411880	the area of national security and hall hazards types of response. And this was a time period that
411880	419560	the FDA was embarking on a medical countermeasures initiative as part of the nation being better
419560	428760	prepared to deal with all different types of hazards. And the FDA's role in that is in the
429620	434360	regulation, the review, regulation, evaluation of what are called medical countermeasures
434360	440880	for biologics, for chemical, for rad nuke related types of events. Because of my burn background,
440880	446480	I was brought in to this effort and introduced to it on the rad nuke side because of the burn
446480	454320	injury nexus. But this opened up a whole new vista for me as far as the ability to kind of
454320	459620	work across agencies, work across other parts of government and exposed me to
461160	468900	areas or fields and disciplines that were really outside of what I trained in. And I found it
468900	476680	fascinating. I found that really being able to work across different parts of government, whether
476680	480960	it's within the Department of Health and Human Services with their Assistant Secretary of
480960	487640	Preparedness and Response and CDC and BARDA, whether it's with DOD assets, whether it was with
488380	495300	other, again, other parts of the government Department of Homeland Security. This to me
495300	502620	was extremely exciting and intriguing because I felt that we have not only an important role to
502620	509220	play and that there is a value that we can add here in a manner that hadn't been really quite
509220	517040	captured. For me, this was more fascinating than simply doing reviews of pre-market applications
517040	524140	or submissions, although I enjoyed seeing that and being closely involved in seeing
524140	531140	or being directly responsible for new technologies coming on the market. This was just a more
531140	539660	100,000 foot, 200,000 foot view of what's happening within the entire kind of healthcare and public
539660	545040	health, critical infrastructure is what we call it. One of the 16 sectors of critical infrastructure
545040	551580	is healthcare and public health. So when an opportunity arose for me at the center where I am,
551580	558340	for devices and radiological health, to assume the role of director of emergency preparedness
558340	563300	operations, medical countermeasures, I applied for that, I got into that, and it was within
563300	571740	that role that we were dealing with, again, all hazards types of issues, preparedness and response.
572260	579340	And I probably was in that role for maybe no more than six months, six or eight months, when
579340	589160	we were faced with an inquiry that had come in. Actually, it was essentially a delivery of a
589160	597200	package of vulnerabilities, security vulnerabilities of medical devices that was brought forward to us
597340	603960	by two very prominent security researchers who worked in medical devices. And between
603960	610320	their kind of providing this to our center, along with a communication that came separately from
610320	616680	Department of Homeland Security's, what was called then ICS-CERT, the Industrial Control System
616680	622940	CERT, Cyber Emergency Response Team, this was rather, this blindsided us, I will be very honest
622940	630220	with you, because we had never dealt with that before. And we had to quickly figure out what
630220	635000	does this actually mean? What do these, you know, bring 400 vulnerabilities to our attention
635580	641060	across all these medical devices? We weren't certain even how to interpret that, and what
641060	649980	does that mean as far as next steps, action taken? It was a decision made by FDA leadership at the
649980	656520	time, because this was part of a response role, a response responsibility at the time,
656520	664040	that this would reside in my shop. Did you want to ask something? Yeah, that was 2013.
665920	674800	So, the thought was, we already had demonstrated with respect to other hazards and other public
674800	679900	health emergencies around medical devices that were brought to our attention, good processes
679900	687700	and capabilities, that this should also reside in our shop. But I will tell you that I had a
687700	692420	small team, this our Emergency Medical Countermeasures Team, of individuals who really
692420	700260	had no, also like myself, no visibility on this area. Again, I'm a burn surgeon, okay, doing other
700260	705700	things, but certainly not a cybersecurity subject matter expert, or anyone who really had any
705700	714940	interfacing with security engineering. And what we needed to do was very quickly mobilize,
714940	722920	mobilize a team, mobilize working group across the center of experts that really took us through
722920	729040	the equivalent of the total product lifecycle. So, folks who were involved in terms of understanding
729040	736080	security engineering and software, pre-market as well as post-market surveillance, compliance,
736080	744080	on the policy side. We have within FDA, surprise, surprise actually, we have an entire Office of
744080	748980	Science and Engineering Laboratory, which is involved in what's called regulatory science
748980	755320	research. So, the researchers, the staff there who are researchers, many of them have these
755320	763980	capabilities too. And we quickly established this working group, and by necessity worked together
763980	773240	with DHS ICS CERT in partnership, and had to come to terms with what do these vulnerabilities mean,
773240	777780	what are the actions that we're going to take with manufacturers, and how are we going to come to
777780	785160	some resolution on that. This was again, spring-summer 2013. The time frame is important
785160	792840	because this was during the Obama administration. In February 2013, there was an executive order
792840	799840	that the Obama administration had issued around the notion of strengthening cybersecurity
799840	805440	across critical infrastructure, public and private sector working together. There was a lot of
805440	811680	emphasis on partnership between the public and private sector. There was a lot of emphasis on
812540	820000	also the role that NIST would have. They were mandated at that time to put together through a
820000	825640	public-private effort, a framework, a framework for cybersecurity, strengthening critical
825640	830180	infrastructure, and they were going to convene workshops around the country. And then DHS at
830180	837280	time was also given several mandates through this executive order to undertake, to pull together
837280	842020	from across government, as well as to the private sector, different work groups that were going to
842020	847500	be assigned different functions. And so when this hit us, one of the first things that we also
847500	852340	recognized is we need to throw ourselves in here. We need to roll up our sleeves, raise our hands to
852340	857240	volunteer and say, this is the way that we are going to learn. This is the way that we are going
857240	864300	to understand our landscape best and learn through doing and partnering and collaborating.
864820	872560	And that kind of brought us to, over time, the recognition that there's, you know, there's
872560	877380	traditional stakeholders in our ecosystem, and then there were new, non-traditional ones who
877380	883560	we had never interacted with before. Security researchers, white hat hackers, whatever one
883560	890460	wants to term them, you know, our feeling was this is a community with whom we've not engaged.
890960	897080	They don't know us. We don't know them. There's a necessity to work together. The value that
897800	904240	hackers bring to the medical device space is enormous. And I emphasize that because
904240	912140	medical device manufacturers have the capability of being able to design devices with safety in
912140	918100	mind, with safety and effective performance, but the notion of building security in by design
918100	923960	was something kind of a very new and foreign concept. And here we were going to need to
923960	929960	really shift the culture and the way of thinking around new devices coming on the market.
930820	938740	So we probably released, the pre-market guidance draft got written and released
938740	944960	in literally like six weeks. That is a just something that was a miracle that like never
944960	951660	happens, but it was recognition that policy has to be, the policy needed to be in place.
951960	959280	And we subsequently finalized it over the course of the year and convened our first public
959280	964800	workshop in 2014 around medical device cybersecurity. And it's an interesting
964800	971840	workshop if you were to go back and watch the webcast or listen to the, listen to it or look
971840	979380	at the archives, because there are a lot of the stakeholders in the room who viewed themselves
979380	988560	somewhat adversarially in years past. Security researchers, manufacturers, healthcare organizations,
988560	996180	other government agencies. And the idea of this was to really foster open dialogue and to allow
996180	1002280	for the pain points that different stakeholder groups are experiencing for them to be
1002280	1008040	vocalized and to not necessarily come up with solutions. We weren't going to have that happen
1008040	1016320	in two days, but we had to give voice to everyone who deserves a seat around the table. And that
1016320	1024840	has been a critical principle that we continue to amplify over the years, this
1026440	1034960	importance of hearing from every stakeholder that's involved and recognizing the value that
1034960	1041620	different stakeholder groups bring, particularly because we know that cybersecurity within this
1041620	1047020	space, as in other spaces, is one of really that shared ownership or shared responsibility that we
1047020	1052380	talk a lot about. So we've emphasized whole of community, not just whole of government,
1052860	1060040	began by really reaching out to security researchers. I Am the Cavalry has been,
1060040	1068420	had a major, major impact on the work that we've done in these years, you know, really back from
1068420	1078820	2013-14 to the present. And I can't even kind of quantify, you know, what extraordinary benefit
1078820	1085760	we've derived from that relationship that we have with I Am the Cavalry and the role that they have,
1085760	1094300	not only in having educated us around the challenges that are faced in the hacker universe,
1094300	1104720	but also an opportunity for them to help inform us with respect to policy issues and areas in other
1105340	1112520	sectors, in other industry verticals that we are able to learn from and to take those experiences
1112520	1118720	and apply them within the healthcare and medical device space. How did the approach to the post
1118720	1124120	market guidance that was really subsequently changed from the way that the pre-market guidance?
1124120	1131600	So very thoughtful and deliberate and intentional. So while we were kind of going through all this,
1131600	1136940	we also are continuing, let's remember to get vulnerabilities brought to our attention. Things
1136940	1144660	just don't stop. And we have to think about what, how are we handling these vulnerabilities? How are
1144660	1149780	we managing them? What are our expectations of manufacturers? Because we had already established
1150040	1157700	a lot in the way of a network of friends. It's like call a friend, right? You know, being able to
1158340	1166340	have a very broad network of subject matter experts to help support and inform the policy
1166340	1173900	approach we were going to take was very, very different than where we were a year and a half,
1173900	1183180	two years prior. And so while we issued the post market initially as draft in early 2016,
1183180	1190700	we'd been already working and socializing the concepts and a framework behind it for at least
1190700	1199920	the 12 months prior. And that was critical because the way FDA works with regard to guidances, if
1199920	1205820	we'll issue something as draft, there's an open comment period, a public docket that is opened.
1205820	1210700	It's generally a common period of around 90 days. Sometimes it's extended for longer than that.
1210700	1216980	And then we'll look at those comments, digest them and determine in finalizing the guidance,
1216980	1222880	what is appropriate, what we think is going to make for a more robust guidance, one that is
1222880	1227720	implementable. The implementable piece is also critical and that plays a very important role
1227720	1233120	with the post market guidance. So the fact that we had done all of this kind of socializing and
1233120	1239580	homework in advance, made it much easier to finalize this guidance. It was issued in January
1239580	1246840	2016. It was finalized by December 2016. Also a very, very quick turnaround for government within
1247400	1254920	a year to go from draft to final. But it's because we had already addressed a fair amount of
1254920	1261340	what we thought would be feasible and what we thought would be the right approach to take.
1261340	1267040	But importantly, there were certain ambitious goals that we had set forth in the draft that
1267800	1273340	the hacker community had come to us and said, based upon what's observed in other industrial
1273340	1280480	control system sectors, don't think that you're going to get that. You may want to think about
1280480	1285140	the crawl, walk, run approach. You may want to think about while you're really bringing
1285140	1293780	industry up to speed here and raising the bar, that's done more incrementally. And some of
1293780	1299320	the expectations that might've been set in the draft may have been a bit too ambitious. And we
1299320	1307760	took that in mind and did further tweaking on it and have since reissued it. I think we still have
1307760	1316720	an ambitious goal there, but we're seeing that manufacturers, for the most part, are striving
1316720	1323140	to attain that goal. The other piece of it that was important is to build in a regulatory incentive.
1323140	1327760	So that was another element of the post-market guidance that we thought was really very,
1327760	1333740	very important, that it be clearly articulated some of the differences that you have
1334580	1341640	in traditional medical device-related management and what our expectations were with cybersecurity,
1341640	1349260	that manufacturers be far more nimble, that they accelerate the type of assessment
1350120	1356240	of vulnerabilities, validating them and remediation of those vulnerabilities. And they do that in a
1356240	1361960	much more accelerated timeframe than what would otherwise perhaps happen with other types of
1361960	1368960	device-related malfunctions. And so that was one of the reasons why we built in certain incentives
1368960	1375440	to try to push or drive for a faster type of response, which, as you can imagine,
1375440	1390320	in cybersecurity becomes really critical to do. The question was, I had made the distinction that
1390320	1396240	in the draft guidance, we put a very, very ambitious timeframe down. In the final guidance,
1396240	1402000	we pulled it back a little bit. So if I'm remembering correctly, and Seth is here as well,
1402000	1409760	in the draft, I think that we had only inserted kind of a sense of a 30-day timeframe all around
1409760	1415440	for really being able to, once the vulnerability was discovered and reported to the manufacturer
1415440	1421400	at being time zero, they really needed to be able to do a full assessment, validation,
1421400	1428620	determine what needs to be done in order to reduce the risk of that vulnerability and communicate it
1428620	1436320	out and take that action in a 30-day timeframe. So that's pretty aggressive. What we did was,
1436320	1444060	we still wanted to kind of demonstrate and really push industry to understand that
1444060	1449000	what we're talking about where security and safety are crossing over, and safety issues
1449000	1455860	are really critical to us here. So we offered, we provided a 30, what we call the 30-60-day
1455860	1463480	timeframe. So 30 days being from time zero, again, from when awareness of the vulnerability going
1463480	1469760	through the assessment and all of that process, with the expectation that manufacturer communicate
1469760	1476500	by the end of 30 days around this vulnerability, what its potential impact is, and at minimum,
1476500	1483060	what compensating controls or temporizing measures would be available in order to reduce risk.
1483060	1489160	And then in the following 30 days, so that would be 60 days total, that a more definitive
1489760	1495280	fix would be available for deployment and subsequent communication with that.
1495280	1502320	So that was the more measured, if you will, or more modified timeframe. Yeah.
1503180	1509620	So the post-market guidance comes out, and there have subsequently been actions taken by the FDA,
1509620	1513000	including some recalls. Can you talk to us about some of those recalls?
1513000	1519800	Yes. So I referenced the fact that we built in a regulatory incentive. So I need to just explain
1519800	1525900	what that incentive is to explain what the recalls are about. The regulatory incentive is to kind of
1525900	1532400	drive towards a compressed timeframe. And we put in three criteria within the post-market guidance
1532400	1539260	that if you meet that timeframe, that you, and that secondly, that there had been no adverse events,
1539260	1544760	no injuries or fatalities as a result of this vulnerability, therefore it's not been exploited
1544760	1551320	to the manufacturer's awareness. And the third one was the manufacturer must be an active participant
1551320	1558100	in what we call a medical device ISAL, vulnerability ISAL, information sharing analysis organization.
1558100	1563100	That parenthetically, the ISAL concept also came out of the Obama administration. It was through
1563100	1570160	an executive order that ISALs kind of form as kind of small satellite communities where there would
1570160	1576140	be information sharing across those. And we recognize the criticality of information sharing
1576140	1583920	within and across industry as a means of being able to disseminate information more rapidly around
1583920	1589600	vulnerabilities that may not necessarily impact one manufacturer, but may impact many to really
1589600	1596420	try to, again, institute a more proactive posture. So those are the three criteria that
1596980	1603480	we described in the post-market guidance. Manufacturer, if you meet those three criteria,
1603480	1610040	then we will not enforce what's called Part 806 corrections and removals, which is essentially
1610360	1616220	a voluntary recall, okay, that we would use enforcement discretion. And for a manufacturer,
1616220	1620460	that would be a good thing because going through a voluntary recall process, you know, brings a lot
1620460	1625480	of attention. There's a lot of administrative stuff that goes along with that. So if there's
1625480	1632900	another way to do that, which is going to accelerate a fix and alleviate for the manufacturer
1632900	1640380	the need to go through this formal process, that would be good. At the same time, what's therefore
1640380	1645720	implied and actually stated within the guidance, if you're not able to meet all three criteria,
1645720	1651140	and we provided examples of that in an appendix, well, then you're, you know, obligated to go
1651140	1656120	through the normal Part 806 corrections and removals. Then you're going through a voluntary
1656120	1663000	recall process. And we recognized dealing with a rather immature ecosystem anyway,
1663000	1668240	and some very difficult challenges with some medical devices, that 60 days is likely not
1668240	1675300	going to be enough for certain devices to have permanent fixes associated with them.
1675300	1682920	So by nature, we knew that there are going to continue to be recalls over time, because
1682920	1689820	it's just going to be a difficulty for manufacturers of certain types of devices to be
1689820	1696980	able to not only do that full validation and assessment, but to actually be able to create
1696980	1702640	an appropriate update or patch or fix that has to also be tested, right? It's got to be tested to
1702640	1708980	make sure that it's not going to interfere with the device's safe performance. So that within 60
1708980	1715200	days may represent an impossibility in certain situations. And in those cases, a manufacturer
1715200	1733050	is going to need to undergo a full Part 806 with a recall. So I can't quote you a percentage. What
1733050	1739050	I would say to you is the following from anecdotal from our experiences thus far. And that is that
1739790	1746130	well, there are a few conditions here to talk about. Number one, the implantable devices
1746130	1753810	represent the hardest fixes. So you want to think about the pacemakers, the ICDs,
1753810	1759830	you know, the implanted defibrillators, neuromodulators, deep brain stimulators,
1759830	1767970	things that are implanted are going to represent a much more complex fix and deployment.
1768670	1775030	So now they don't represent a huge percentage as we were looking across the entire medical device
1775030	1782310	portfolio of devices that we regulate. That's not a, you know, huge chunk. But
1783250	1788970	you have that as one consideration. The other important consideration, which is relevant to
1788970	1795510	this discussion also, is that we've seen a fair amount, particularly with these are high
1795510	1802610	consequence types of vulnerabilities, what can be debate or disagreement across researchers and
1802610	1807750	manufacturers, or researchers, manufacturers, and government with regard to what we like to
1807750	1814790	call ground truth. Is this vulnerability, first of all, real in the sense of does this exist?
1814930	1821110	Has there been an ability to validate its presence? And more importantly, or by taking
1821110	1827150	it a step further, the impact of that vulnerability and the severity of that impact,
1827150	1835170	where its capability of introducing patient harm that we call uncontrolled risk,
1835170	1840360	or risk that we would consider to be not tolerable from the agency's perspective,
1840630	1846110	not tolerable in the way it exists, and therefore requires manufacturer to do something
1846110	1853510	with that device to reduce that risk to an acceptable level. So when you have
1854350	1860650	disagreement across parties with regard to how to assess the assessment of that vulnerability
1860650	1866910	and its impact, that already is going to prolong the timeframe out of the 60 days.
1866910	1874230	So that's the other piece also that has resulted in some of the recalls that have been public,
1874230	1880470	getting to that ground truth has been a very complicated process. And because it's well
1880470	1886790	exceeded 60 days, it's been months in some cases, we've had a few that have gone a year and a half
1886790	1894530	or longer, that has necessitated a recall. And so now we have draft pre-market guidance
1894530	1899230	as an update. Can you tell us a little bit about that draft and what the next steps will be?
1899230	1906570	Yeah. So one of the also important principles where we started out from is that with cyber
1906570	1911910	security really rapidly evolving and knowing where we and the ecosystem were starting,
1911910	1916990	we recognized that this is an area that we were going to need to iterate on, on a continuum.
1916990	1922310	And then what we issued back a few years ago, that's fine for a foundation,
1922310	1928290	as appropriate as a baseline kind of setting. But given everything that we have observed,
1928290	1932810	everything that we have learned and that we've experienced over the past few years,
1932810	1937090	whether it's WannaCry as an example, or NotPetya that was talked about earlier,
1937430	1943870	whether it's vulnerabilities that have been brought to our attention by security researchers,
1943870	1948830	we recognize through those learnings that, yeah, we got to be continuing to raise the bar
1948830	1955930	and our expectations of manufacturers are that much more rigorous than they were in 2014.
1955930	1962700	So 2014 sets the floor and the principles upon which we continue to build. Did you have a question?
1983440	1992220	Yes. Yes. And those are examples of where we've had this protracted phase of trying to get to
1992220	1998500	ground truth around vulnerabilities where we have had to insert ourselves as part of that process,
1998500	2003300	working together with the security researchers, working with subject matter experts that are
2003300	2010460	technically capable of being able to further understand exactly what the researchers have
2010460	2019800	been able to demonstrate and working with the manufacturer. And that is a challenging space,
2019800	2030880	that I'll take a detour for a second, is a reason why we put forward in 2018, in the spring,
2031000	2037720	a medical device safety action plan for medical devices in general with an entire section on
2037720	2044400	cybersecurity. And we put forward a proposal there for something that we call the CIMSAB,
2044400	2050740	CyberMed Safety Analysis Board. And the concept behind the CyberMed Safety Analysis Board is just
2050740	2056960	that, recognizing that we are challenged today, we have a gap of knowledge and understanding
2056960	2064840	within the ecosystem where you can have vehement disagreement across different parties as to
2064840	2070240	not just the vulnerability's existence, you know, I think that we can get to that point,
2070240	2076320	but what is the impact of that vulnerability from a clinical perspective and therefore what action
2076870	2083900	would need to be taken. So, having an entity that's more of a neutral body, a neutral entity
2083900	2091700	that's a public-private model that includes technical experts, clinical experts, engineering
2091700	2097620	experts, security experts, that government has a seat at the table as well, but government doesn't
2097620	2106600	own this entity. Think somewhat like akin to an NTSB model, not exactly like that, but the idea,
2106600	2112400	and not waiting for something bad to happen, but rather when a vulnerability of this high consequence
2112400	2119960	is brought to our attention, it's, you know, we find ourselves having to reach out to multiple
2119960	2126180	parties trying to get to this place of understanding really what the impact of that vulnerability is.
2126180	2133380	How much better would it be if there was a pre-positioned body of experts that are
2133380	2140820	able to come together in real time to fully assess through each kind of expert's lens, through
2140820	2146260	different disciplines, really what the impact is of that vulnerability and what are the necessary
2146260	2153520	next steps to take. So, that is something that we are proposing for initial seed funding. We're
2153520	2160900	hoping that we can get some funding for that with the idea of it being subsequently owned and
2160900	2166840	operated by the public-private sector, kind of through a public-private sector model. But going
2166840	2174600	back to the new pre-market update, so in reflecting on what we've learned, we come to three basic
2174600	2180740	principles. We call them trustworthiness, transparency, and resilience, recognizing that
2180740	2184840	we've got to build those foundational principles into everything that we do.
2184840	2189620	It has to start with the pre-market, but it really goes throughout the entire post-market
2189620	2196260	lifespan of a device. However, you're only going to get to that if you build it from the start,
2196260	2205770	from scratch, with all of those concepts in mind. And so, we released that in October. Some of the
2205770	2212470	key elements there that were really important ahas for us, number one, was the concept of bill of
2212470	2221250	materials, right? So, WannaCry had transpired and a recognition that you can only protect what you
2221250	2226610	know that you have. For an end-user organization, for customers, for healthcare organizations
2226610	2232230	that have thousands upon thousands, if not, in some cases, millions of medical devices residing
2232230	2238230	on systems, how can they possibly protect their systems, protect the devices in terms of their
2238230	2245710	performance if they do not know what the composition is, the components are of those
2245710	2249710	devices, and whether there are vulnerabilities that have already been identified within those
2249710	2259150	devices. So, while we are heavily engaged in a fantastic effort, a multi-stakeholder effort
2259150	2265350	through NTIA, Department of Commerce, the SBOM, or the Software Transparency Initiative,
2265350	2271010	which is fantastic. At the time, we also thought we were going to kind of even raise it a bar
2271010	2276030	further and say, well, it's not just software, we're concerned about hardware as well. So, we
2276030	2281810	called it the CBOM in the draft guidance, the Cybersecurity Bill of Materials, with the expectation
2281810	2287230	that manufacturers actually would have to provide that to their customers, not just to us, to their
2287230	2294590	customers. We've had a workshop since, our docket as well, the comments have come in, and there's
2294590	2299830	been recognition across the board by people that are a lot more knowledgeable and experienced than
2300350	2307930	us, that that's great, and that's aspirational, but that's maybe taking it down the road.
2307930	2313250	Really, what you want to start with is, let's get that SBOM, let's get that Software Bill of
2313250	2319290	Materials in place. So, as we're in the process right now of revising and kind of taking into
2319290	2326430	account the comments that came into the guidance draft that we updated, the update that we issued
2326430	2334510	in 2018, we're going to put that back as SBOM, as opposed to CBOM, and once we get there,
2334510	2340090	then we'll move forward to CBOM as we continue to iterate and make things more rigorous.
2340090	2345850	Some of the other pieces that were important there was putting in an expectation that
2345850	2353250	manufacturers have to include threat modeling and documentation of threat modeling in the design
2353250	2360250	of their devices. Third item that was really important was, manufacturers need to include
2360250	2367370	in their pre-market submissions evidence that devices are capable of being updated and patched
2367370	2375170	through the lifetime of their device. We are dealing with the pain of legacy devices right now.
2375490	2384210	That has been an area that is going to be very difficult to address over the next few years.
2384210	2389810	What we can do while we're trying to address that through one arm, certainly with new devices
2389810	2397050	coming on the market, the expectation that they have greater resilience, that they are capable
2397050	2403230	of being updated and patched. We want to see that as part of the pre-market submission, not to find
2403230	2409690	out once a device is in use that this device is brittle and any attempt to trying to update it is
2409690	2416430	actually going to knock over the device. So those are some of the concepts that we incorporated
2416430	2424610	within the new pre-market guidance. And we're, again, once more in the process of revising it
2424610	2431310	right now with the hopes of being able to release it again by the end of this calendar year.
2432070	2436810	Okay, and to maximize time for questions, I will open it up. Are there any other questions
2436810	2445850	that we can engage on? Could you use the mic? Thanks.
2464580	2468560	Yeah, really good question. I think right now we are also
2470120	2479900	dealing with the reality at FDA that we are responsible for making changes that were
2479900	2487220	mandated by 21st Century Cures Act as well, which brings into account a lot of interesting
2487220	2495120	questions around products that have some functions that are considered medical device functions,
2495120	2502860	upon which we have direct authorities and some functions now which may not actually be considered
2502860	2510260	medical devices. So you can imagine how sort of convoluted that can be as far as how to approach
2510260	2519060	that from a cybersecurity perspective. Trying to create an artificial walling parts of it off
2519060	2524280	from what we would look at versus what we wouldn't look at is very complex. We're working through
2524280	2530260	some of those challenges now, but what I would say is the following, and I had this aha moment
2530260	2535200	actually when I was sitting and listening to Alan Friedman the other day at B-Sides, you know, talk
2535200	2541940	about the Software Transparency Initiative, the effort, and the group's work in terms of defining,
2541940	2547540	well, what exactly is SBOM? What really is it? And he said where the group had come to was
2547540	2554020	X includes Y. Let's not talk about functions right now or, you know, let's talk about what's
2554020	2560760	included within that. And for me, that really resonated because that meant, you know, the SBOM
2560760	2566160	is really intended to be that protective measure, that ability to mitigate for the end user, for the
2566160	2573560	customer. And so trying to separate out or divorce out pieces of it that would not necessarily be
2573560	2581620	under our authority doesn't make sense. As we look towards SBOM, we would be looking at the product
2581620	2590400	in its totality, even those components of the product that may not necessarily have a direct
2590400	2596140	medical device performance function, but that wouldn't matter to the end user, right? Their
2596140	2602140	concern is for the vulnerability and their exposure. So that's, you know, that's kind of
2602140	2612620	how we're thinking about it right now. I was curious about the scope of the vulnerabilities
2612620	2618700	that your team is responsible for addressing. Is it just physical hazards or the function
2618700	2623420	of the hardware device or do you also look at things like user privacy, for example, or other?
2623440	2628860	Yeah, that's a really great question. I didn't cover that at all. So the interesting thing is
2628860	2634500	that across the Department of Health and Human Services, different agencies, different operating
2634500	2643040	divisions have different authorities that they are responsible for. ONC, I'm sorry, not ONC, OCR,
2643040	2650560	Office of Civil Rights within HHS, is responsible for HIPAA privacy and security. That's not under
2650560	2660960	our remit. And so while we would not minimize the importance of privacy, of patient information,
2660960	2668720	clearly that's very important. Our authorities and our mission is specific to the safe and
2668720	2674300	effective performance of devices and protecting patients from physical harm if the device
2674300	2682420	was not performing properly, it was not performing at all. So what we did in guidance that we've
2682420	2689780	previously issued, though, is we state very clearly, manufacturers, you are still under other
2689780	2696420	obligations that you need to abide by, including HIPAA privacy and security. Here's what we're
2696420	2702720	doing to make it easier for you to abide by that. So we have something called controlled
2702720	2708400	vulnerabilities, vulnerabilities that in their assessment don't necessarily result in a patient
2708400	2715300	harm. What we've said there is that changes that are made solely to strengthen cybersecurity
2716260	2722960	and data privacy-related issues would fall in that category. They are considered,
2722960	2729020	from our perspective, routine updates and patches. They do not come to the agency for re-review.
2729020	2734660	Manufacturers, you should just go and do that. You should just take those proactive steps to
2734660	2741860	make sure that you can then be in compliance with the other regulations under other parts
2741860	2747360	of government that you are required to abide by. And we're not going to impede that process.
2747360	2751940	We're not going to delay that because we say we need to see those. We do not need to see those.
2755420	2763980	Other questions? What do you foresee to be the biggest challenge going forward in the
2763980	2771420	next five years for the FDA to the point of security? Biggest challenge? There's a few of
2771420	2781280	them. Some of them are wrapped together. I think holistically across the ecosystem,
2781280	2786860	so it's not just FDA's challenge and this is where we really have to bring other parties together,
2786860	2795760	it is the legacy system and the legacy device issue. And I mentioned that because of the
2795760	2804880	economics associated with that as well and the recognition that while new devices can be made
2804880	2813760	available to healthcare organizations, healthcare organizations are not necessarily well-poised
2813760	2820240	financially or through their business models to be able to take in those new devices in order to
2820240	2829940	get rid of the old legacy drag. And so you're left with certain systems that are not capable
2829940	2835700	of being updated and patched and you're left therefore with certain exposures and vulnerabilities
2835700	2845420	and recognizing that a cyber campaign can occur at any time and can exploit these vulnerabilities
2845990	2854680	that leaves our critical infrastructure vulnerable. So how do we address that from
2854680	2860940	multiple facets? Recognizing that, yeah, we have some tools in our back pocket that we could pull
2860940	2866000	out. Some of them are a little bit drastic that we would rather not move to. We'd like to figure
2866000	2870960	out through efforts of the public and private sector working together and there are a lot of
2870960	2878040	different consortia doing that under even what's called the big umbrella, the Healthcare Sector
2878040	2886680	Coordinating Council to really analyze some of the economics here and then what are the options,
2886680	2893400	what might be the incentives that could be put forward to get these really older devices out of
2893400	2900240	use and bring newer devices that are more rigorous and more secure in place. I don't know, Seth, if
2900240	2905100	there's anything that you want to add, but that's, you know, we've got lots and lots of challenges
2905100	2910240	of different scale, but when I think of something that's really big scale that's going to require
2911320	2915980	a whole universe of effort, that's probably one of the biggest ones.
2916680	2922440	And with that, we will have to wrap things up. Let's all thank Dr. Suzanne Schwartz for joining