WEBVTT

00:00.000 --> 00:01.420
Thank you all for being here.

00:01.420 --> 00:03.060
I'm Andrea Matwishan.

00:03.060 --> 00:11.680
I'm a professor at Penn State, and it is my distinct pleasure to share the next hour with Dr.

00:11.680 --> 00:14.140
Suzanne Schwartz of the FDA.

00:15.060 --> 00:22.460
And I realize I owe Anthony Ferrante a mug, because this is a... I promised everyone coffee time.

00:22.460 --> 00:27.020
There's a de minimis mug that shall be yours.

00:27.020 --> 00:29.740
Maybe when Q&A starts, I'll produce the mug.

00:29.740 --> 00:30.080
Okay.

00:30.080 --> 00:31.820
Like habeas corpus only.

00:31.820 --> 00:32.960
Habeas mugus.

00:32.960 --> 00:33.400
I don't know.

00:33.400 --> 00:33.760
Whatever.

00:33.760 --> 00:53.860
Anyway, so today we'll discuss with Suzanne some of the current FDA approaches to security and learn a little bit about where we see things going with respect to medical devices and security.

00:53.860 --> 00:59.480
But first, I'd love to start with just asking Suzanne, how did you become interested in security?

00:59.480 --> 01:02.600
You weren't born doing your role.

01:02.600 --> 01:04.260
You arrived in your role.

01:04.260 --> 01:18.540
So tell us a little bit about your history and how you went from prior iterations of self to your current job and your deep engagement with this area.

01:18.540 --> 01:24.700
And I will say, without hesitation, the FDA is probably the most active agency right now on security.

01:24.700 --> 01:28.380
And kudos to the agency on that.

01:28.380 --> 01:35.700
And you're looking at one of the main reasons for the FDA's deep interest and forward-looking approach in security.

01:35.700 --> 01:40.740
So I'd love to hear the story of how you became yourself.

01:41.080 --> 01:41.480
Thank you.

01:41.480 --> 01:44.740
And I have to say that, no, I wasn't born into security.

01:44.740 --> 01:48.340
I think when I was born, computer security actually didn't exist.

01:48.340 --> 01:50.640
I'm dating myself here.

01:51.200 --> 02:07.300
In any case, I'm a physician and my background and training actually, clinically, is in the area of surgery, general surgery , specifically burn injuries and trauma, which is far afield from the area of cybersecurity.

02:07.640 --> 02:13.660
I've done a lot of work as both an academic, what's called a surgeon-scientist.

02:13.660 --> 02:32.200
I spent quite a bit of time actually within an academic clinical center in New York City, Weill Cornell Medical Center, which is part of kind of the New York Presbyterian Complex, doing both burn-related and general wound repair-related clinical care and research.

02:32.200 --> 02:52.540
And one thing I would say just about myself is the desire to pursue research and innovation and looking at new approaches, new technologies that can further benefit patients.

02:52.540 --> 02:56.000
And that's a common thread really throughout my entire career.

02:56.000 --> 03:04.840
Aside from the work that I had done on the clinical side in the hospital setting, I've also been on the industry side.

03:04.840 --> 03:23.160
Before coming to the FDA, I worked in the private sector as a medical director of a startup company, a startup that was involved in actually producing and testing a tissue-engineered skin, skin for burn injuries and wound repair.

03:23.160 --> 03:43.640
And it was just kind of over the next few years, this kind of realization that for me, being able to straddle multiple sectors of the universe, the research and science side and technology and understanding of the clinical needs, a desire to see technologies get to the bedside.

03:43.640 --> 03:54.480
And in that role as a medical director, I did a lot of interfacing with the FDA in order to get clinical trials through and to be part of putting applications together.

03:54.480 --> 04:04.100
So there's a certain nexus there that for me was very appealing in terms of pursuit of what can I do to be more impactful?

04:04.100 --> 04:15.920
What can I do from a more general, broader public health perspective in seeing patients get new therapies?

04:16.100 --> 04:24.840
And so fast forward a bit in 2010, I came to the FDA in a role that's called the Commissioner's Fellowship.

04:24.840 --> 04:31.080
It's actually unfortunately been phased out or it's in the process of being rebirthed as something a little bit different.

04:31.080 --> 04:47.600
But the idea of a FDA Commissioner's Fellow, we came in as a cohort, were to really bring in individuals with varying subject matter expertise in different disciplines that may be lacking at the agency.

04:47.600 --> 05:03.080
And by bringing in individuals with different subject matter expertise, certainly that benefits the agency by helping us with some of these newer, more advanced methodologies and technologies to help in review of those and an assessment of those.

05:03.080 --> 05:14.460
And at the same time, the fellowship would enable the individual to be exposed to the regulatory framework and what's happening with the FDA.

05:14.460 --> 05:34.180
So, whether that individual then ends up going back out to the private sector and acting or serving as an ambassador to the environment in which they're going to work, where they understand how regulation in the medical product area works, that's one benefit.

05:34.180 --> 05:40.580
The other benefit is for those individuals where it was determined to be a really mutually beneficial match.

05:40.580 --> 05:51.260
So, for myself, I was very excited about the work that I was assigned at the FDA when I came as a Commissioner's Fellow.

05:51.260 --> 06:10.060
And I saw a huge opportunity to have impact, very, very challenging and novel emerging areas and an opportunity to learn more and be that continuous constant learner, which is something also that's important to me in whatever I do.

06:10.060 --> 06:21.080
So that at the end of every day, I feel, wow, I've learned something new or been challenged with something different and tomorrow's a new day and who knows what that's going to bring.

06:21.520 --> 06:40.780
In my journey at the FDA, although I came in as a Commissioner's Fellow and then because I'm a physician, that translates to being a medical officer, because I was also representing a field that was a very unique field that you don't have that many surgeons around that are burn surgeons.

06:40.780 --> 06:50.600
It is a small field, but there were connections to the area of national security and hall hazards types of response.

06:50.600 --> 07:03.860
And this was a time period that the FDA was embarking on a medical countermeasures initiative as part of the nation being better prepared to deal with all different types of hazards.

07:03.860 --> 07:19.460
And the FDA's role in that is in the regulation, the review, regulation, evaluation of what are called medical countermeasures for biologics, for chemical, for rad nuke related types of events.

07:19.460 --> 07:28.060
Because of my burn background, I was brought in to this effort and introduced to it on the rad nuke side because of the burn injury nexus.

07:28.460 --> 07:47.560
But this opened up a whole new vista for me as far as the ability to kind of work across agencies, work across other parts of government and exposed me to areas or fields and disciplines that were really outside of what I trained in.

07:47.560 --> 07:49.660
And I found it fascinating.

07:49.660 --> 08:09.120
I found that really being able to work across different parts of government, whether it's within the Department of Health and Human Services with their Assistant Secretary of Preparedness and Response and CDC and BARDA, whether it's with DOD assets, whether it was with other,

08:09.120 --> 08:14.700
again, other parts of the government Department of Homeland Security.

08:14.700 --> 08:30.480
This to me was extremely exciting and intriguing because I felt that we have not only an important role to play and that there is a value that we can add here in a manner that hadn't been really quite captured.

08:31.060 --> 08:49.680
For me, this was more fascinating than simply doing reviews of pre-market applications or submissions, although I enjoyed seeing that and being closely involved in seeing or being directly responsible for new technologies coming on the market.

08:49.680 --> 09:02.440
This was just a more 100,000 foot, 200,000 foot view of what's happening within the entire kind of healthcare and public health, critical infrastructure is what we call it.

09:02.440 --> 09:07.280
One of the 16 sectors of critical infrastructure is healthcare and public health.

09:07.280 --> 09:26.620
So when an opportunity arose for me at the center where I am, for devices and radiological health, to assume the role of director of emergency preparedness operations, medical countermeasures, I applied for that, I got into that, and it was within that role that we were dealing with,

09:26.620 --> 09:31.740
again, all hazards types of issues, preparedness and response.

09:32.260 --> 09:45.060
And I probably was in that role for maybe no more than six months, six or eight months, when we were faced with an inquiry that had come in.

09:45.060 --> 10:02.720
Actually, it was essentially a delivery of a package of vulnerabilities, security vulnerabilities of medical devices that was brought forward to us by two very prominent security researchers who worked in medical devices.

10:02.720 --> 10:20.840
And between their kind of providing this to our center, along with a communication that came separately from Department of Homeland Security's, what was called then ICS-CERT, the Industrial Control System CERT, Cyber Emergency Response Team, this was rather,

10:20.840 --> 10:25.960
this blindsided us, I will be very honest with you, because we had never dealt with that before.

10:26.200 --> 10:31.620
And we had to quickly figure out what does this actually mean?

10:31.620 --> 10:38.060
What do these, you know, bring 400 vulnerabilities to our attention across all these medical devices?

10:38.060 --> 10:44.300
We weren't certain even how to interpret that, and what does that mean as far as next steps, action taken?

10:45.440 --> 11:00.520
It was a decision made by FDA leadership at the time, because this was part of a response role, a response responsibility at the time, that this would reside in my shop.

11:00.560 --> 11:01.700
Did you want to ask something?

11:01.700 --> 11:04.040
Yeah, that was 2013.

11:05.920 --> 11:25.540
So, the thought was, we already had demonstrated with respect to other hazards and other public health emergencies around medical devices that were brought to our attention, good processes and capabilities, that this should also reside in our shop.

11:25.540 --> 11:37.080
But I will tell you that I had a small team, this our Emergency Medical Countermeasures Team, of individuals who really had no, also like myself, no visibility on this area.

11:37.080 --> 11:49.420
Again, I'm a burn surgeon, okay, doing other things , but certainly not a cybersecurity subject matter expert, or anyone who really had any interfacing with security engineering.

11:49.740 --> 12:05.640
And what we needed to do was very quickly mobilize, mobilize a team, mobilize working group across the center of experts that really took us through the equivalent of the total product lifecycle.

12:05.640 --> 12:18.040
So, folks who were involved in terms of understanding security engineering and software, pre-market as well as post-market surveillance, compliance, on the policy side.

12:18.040 --> 12:29.680
We have within FDA, surprise, surprise actually, we have an entire Office of Science and Engineering Laboratory, which is involved in what's called regulatory science research.

12:29.680 --> 12:36.980
So, the researchers, the staff there who are researchers, many of them have these capabilities too.

12:36.980 --> 13:00.020
And we quickly established this working group, and by necessity worked together with DHS ICS CERT in partnership, and had to come to terms with what do these vulnerabilities mean, what are the actions that we're going to take with manufacturers, and how are we going to come to some resolution on that.

13:00.280 --> 13:03.720
This was again, spring-summer 2013.

13:03.720 --> 13:08.700
The time frame is important because this was during the Obama administration.

13:08.700 --> 13:24.600
In February 2013, there was an executive order that the Obama administration had issued around the notion of strengthening cybersecurity across critical infrastructure, public and private sector working together.

13:24.600 --> 13:29.560
There was a lot of emphasis on partnership between the public and private sector.

13:29.560 --> 13:36.120
There was a lot of emphasis on also the role that NIST would have.

13:36.120 --> 13:49.000
They were mandated at that time to put together through a public -private effort, a framework, a framework for cybersecurity, strengthening critical infrastructure , and they were going to convene workshops around the country.

13:49.000 --> 14:03.880
And then DHS at time was also given several mandates through this executive order to undertake, to pull together from across government, as well as to the private sector, different work groups that were going to be assigned different functions.

14:03.980 --> 14:10.100
And so when this hit us, one of the first things that we also recognized is we need to throw ourselves in here.

14:10.100 --> 14:15.940
We need to roll up our sleeves, raise our hands to volunteer and say, this is the way that we are going to learn.

14:15.940 --> 14:24.300
This is the way that we are going to understand our landscape best and learn through doing and partnering and collaborating.

14:24.820 --> 14:39.820
And that kind of brought us to, over time, the recognition that there's, you know, there's traditional stakeholders in our ecosystem, and then there were new, non-traditional ones who we had never interacted with before.

14:39.820 --> 14:50.460
Security researchers, white hat hackers, whatever one wants to term them, you know, our feeling was this is a community with whom we've not engaged.

14:50.960 --> 14:52.300
They don't know us.

14:52.300 --> 14:53.380
We don't know them.

14:53.380 --> 14:55.740
There's a necessity to work together.

14:55.740 --> 15:01.840
The value that hackers bring to the medical device space is enormous.

15:01.980 --> 15:22.200
And I emphasize that because medical device manufacturers have the capability of being able to design devices with safety in mind, with safety and effective performance, but the notion of building security in by design was something kind of a very new and foreign concept.

15:22.200 --> 15:29.960
And here we were going to need to really shift the culture and the way of thinking around new devices coming on the market.

15:30.820 --> 15:41.060
So we probably released, the pre-market guidance draft got written and released in literally like six weeks.

15:41.060 --> 15:51.660
That is a just something that was a miracle that like never happens, but it was recognition that policy has to be, the policy needed to be in place.

15:51.960 --> 16:03.860
And we subsequently finalized it over the course of the year and convened our first public workshop in 2014 around medical device cybersecurity.

16:03.860 --> 16:23.820
And it's an interesting workshop if you were to go back and watch the webcast or listen to the, listen to it or look at the archives, because there are a lot of the stakeholders in the room who viewed themselves somewhat adversarially in years past.

16:23.820 --> 16:30.800
Security researchers, manufacturers, healthcare organizations, other government agencies.

16:30.800 --> 16:46.880
And the idea of this was to really foster open dialogue and to allow for the pain points that different stakeholder groups are experiencing for them to be vocalized and to not necessarily come up with solutions.

16:46.880 --> 16:55.440
We weren't going to have that happen in two days, but we had to give voice to everyone who deserves a seat around the table.

16:55.440 --> 17:22.080
And that has been a critical principle that we continue to amplify over the years, this importance of hearing from every stakeholder that's involved and recognizing the value that different stakeholder groups bring, particularly because we know that cybersecurity within this space,

17:22.080 --> 17:28.540
as in other spaces, is one of really that shared ownership or shared responsibility that we talk a lot about.

17:28.720 --> 17:37.340
So we've emphasized whole of community, not just whole of government, began by really reaching out to security researchers.

17:37.340 --> 17:51.000
I Am the Cavalry has been, had a major, major impact on the work that we've done in these years, you know, really back from 2013-14 to the present.

17:51.180 --> 18:14.300
And I can't even kind of quantify, you know, what extraordinary benefit we've derived from that relationship that we have with I Am the Cavalry and the role that they have, not only in having educated us around the challenges that are faced in the hacker universe,

18:14.300 --> 18:36.860
but also an opportunity for them to help inform us with respect to policy issues and areas in other sectors, in other industry verticals that we are able to learn from and to take those experiences and apply them within the healthcare and medical device space.

18:37.180 --> 18:44.120
How did the approach to the post market guidance that was really subsequently changed from the way that the pre-market guidance?

18:44.120 --> 18:48.000
So very thoughtful and deliberate and intentional.

18:48.000 --> 18:56.740
So while we were kind of going through all this, we also are continuing, let's remember to get vulnerabilities brought to our attention.

18:56.740 --> 18:58.600
Things just don't stop.

18:58.860 --> 19:04.300
And we have to think about what, how are we handling these vulnerabilities?

19:04.300 --> 19:05.440
How are we managing them?

19:05.440 --> 19:07.700
What are our expectations of manufacturers?

19:07.820 --> 19:13.380
Because we had already established a lot in the way of a network of friends.

19:13.860 --> 19:15.720
It's like call a friend, right?

19:16.140 --> 19:36.320
You know, being able to have a very broad network of subject matter experts to help support and inform the policy approach we were going to take was very, very different than where we were a year and a half, two years prior.

19:36.640 --> 19:52.760
And so while we issued the post market initially as draft in early 2016, we'd been already working and socializing the concepts and a framework behind it for at least the 12 months prior.

19:53.100 --> 20:05.820
And that was critical because the way FDA works with regard to guidances, if we'll issue something as draft, there's an open comment period, a public docket that is opened.

20:05.820 --> 20:08.540
It's generally a common period of around 90 days.

20:08.540 --> 20:10.700
Sometimes it's extended for longer than that.

20:10.700 --> 20:23.800
And then we'll look at those comments, digest them and determine in finalizing the guidance, what is appropriate, what we think is going to make for a more robust guidance, one that is implementable.

20:23.800 --> 20:29.580
The implementable piece is also critical and that plays a very important role with the post market guidance.

20:29.680 --> 20:37.560
So the fact that we had done all of this kind of socializing and homework in advance, made it much easier to finalize this guidance.

20:37.560 --> 20:40.400
It was issued in January 2016.

20:40.400 --> 20:43.320
It was finalized by December 2016.

20:43.320 --> 20:50.080
Also a very, very quick turnaround for government within a year to go from draft to final.

20:50.080 --> 21:01.340
But it's because we had already addressed a fair amount of what we thought would be feasible and what we thought would be the right approach to take.

21:01.340 --> 21:18.400
But importantly, there were certain ambitious goals that we had set forth in the draft that the hacker community had come to us and said, based upon what's observed in other industrial control system sectors, don't think that you're going to get that.

21:18.860 --> 21:22.400
You may want to think about the crawl, walk, run approach.

21:22.400 --> 21:32.060
You may want to think about while you're really bringing industry up to speed here and raising the bar, that's done more incrementally.

21:32.060 --> 21:38.840
And some of the expectations that might've been set in the draft may have been a bit too ambitious.

21:38.840 --> 21:46.780
And we took that in mind and did further tweaking on it and have since reissued it.

21:46.780 --> 21:58.720
I think we still have an ambitious goal there, but we're seeing that manufacturers, for the most part, are striving to attain that goal.

21:58.720 --> 22:03.140
The other piece of it that was important is to build in a regulatory incentive.

22:03.140 --> 22:21.640
So that was another element of the post-market guidance that we thought was really very, very important, that it be clearly articulated some of the differences that you have in traditional medical device-related management and what our expectations were with cybersecurity,

22:21.640 --> 22:34.740
that manufacturers be far more nimble, that they accelerate the type of assessment of vulnerabilities, validating them and remediation of those vulnerabilities.

22:34.740 --> 22:44.060
And they do that in a much more accelerated timeframe than what would otherwise perhaps happen with other types of device-related malfunctions.

22:44.060 --> 22:58.540
And so that was one of the reasons why we built in certain incentives to try to push or drive for a faster type of response, which, as you can imagine, in cybersecurity becomes really critical to do.

23:04.140 --> 23:14.500
The question was, I had made the distinction that in the draft guidance, we put a very, very ambitious timeframe down.

23:14.500 --> 23:18.520
In the final guidance, we pulled it back a little bit.

23:18.520 --> 23:37.220
So if I'm remembering correctly, and Seth is here as well, in the draft, I think that we had only inserted kind of a sense of a 30-day timeframe all around for really being able to, once the vulnerability was discovered and reported to the manufacturer at being time zero,

23:37.220 --> 23:52.020
they really needed to be able to do a full assessment, validation, determine what needs to be done in order to reduce the risk of that vulnerability and communicate it out and take that action in a 30-day timeframe.

23:52.040 --> 23:54.900
So that's pretty aggressive.

23:54.900 --> 24:11.180
What we did was, we still wanted to kind of demonstrate and really push industry to understand that what we're talking about where security and safety are crossing over, and safety issues are really critical to us here.

24:11.180 --> 24:16.900
So we offered, we provided a 30, what we call the 30-60-day timeframe.

24:16.900 --> 24:35.220
So 30 days being from time zero, again, from when awareness of the vulnerability going through the assessment and all of that process, with the expectation that manufacturer communicate by the end of 30 days around this vulnerability, what its potential impact is,

24:35.220 --> 24:43.060
and at minimum, what compensating controls or temporizing measures would be available in order to reduce risk.

24:43.060 --> 24:55.280
And then in the following 30 days, so that would be 60 days total, that a more definitive fix would be available for deployment and subsequent communication with that.

24:55.280 --> 25:01.440
So that was the more measured, if you will, or more modified timeframe.

25:01.800 --> 25:02.320
Yeah.

25:03.180 --> 25:10.840
So the post-market guidance comes out, and there have subsequently been actions taken by the FDA, including some recalls.

25:10.840 --> 25:13.000
Can you talk to us about some of those recalls?

25:13.000 --> 25:13.100
Yes .

25:13.480 --> 25:18.800
So I referenced the fact that we built in a regulatory incentive.

25:18.800 --> 25:23.700
So I need to just explain what that incentive is to explain what the recalls are about.

25:23.740 --> 25:28.460
The regulatory incentive is to kind of drive towards a compressed timeframe.

25:28.480 --> 25:46.860
And we put in three criteria within the post-market guidance that if you meet that timeframe, that you, and that secondly, that there had been no adverse events, no injuries or fatalities as a result of this vulnerability, therefore it's not been exploited to the manufacturer's awareness.

25:46.860 --> 25:58.100
And the third one was the manufacturer must be an active participant in what we call a medical device ISAL, vulnerability ISAL, information sharing analysis organization.

25:58.100 --> 26:02.800
That parenthetically, the ISAL concept also came out of the Obama administration.

26:02.800 --> 26:12.620
It was through an executive order that ISALs kind of form as kind of small satellite communities where there would be information sharing across those.

26:12.620 --> 26:30.700
And we recognize the criticality of information sharing within and across industry as a means of being able to disseminate information more rapidly around vulnerabilities that may not necessarily impact one manufacturer, but may impact many to really try to,

26:30.700 --> 26:34.180
again, institute a more proactive posture.

26:34.180 --> 26:39.960
So those are the three criteria that we described in the post-market guidance.

26:40.220 --> 26:54.320
Manufacturer, if you meet those three criteria, then we will not enforce what's called Part 806 corrections and removals, which is essentially a voluntary recall, okay, that we would use enforcement discretion.

26:54.660 --> 27:01.520
And for a manufacturer, that would be a good thing because going through a voluntary recall process, you know, brings a lot of attention.

27:01.680 --> 27:05.000
There's a lot of administrative stuff that goes along with that.

27:05.040 --> 27:17.300
So if there's another way to do that, which is going to accelerate a fix and alleviate for the manufacturer the need to go through this formal process, that would be good.

27:17.840 --> 27:34.960
At the same time, what's therefore implied and actually stated within the guidance, if you're not able to meet all three criteria, and we provided examples of that in an appendix, well, then you're, you know, obligated to go through the normal Part 806 corrections and removals.

27:34.960 --> 27:37.720
Then you're going through a voluntary recall process.

27:37.720 --> 27:55.300
And we recognized dealing with a rather immature ecosystem anyway, and some very difficult challenges with some medical devices, that 60 days is likely not going to be enough for certain devices to have permanent fixes associated with them.

27:55.300 --> 28:21.540
So by nature, we knew that there are going to continue to be recalls over time, because it's just going to be a difficulty for manufacturers of certain types of devices to be able to not only do that full validation and assessment, but to actually be able to create an appropriate update or patch or fix that has to also be tested,

28:21.540 --> 28:21.740
right?

28:21.740 --> 28:27.480
It's got to be tested to make sure that it's not going to interfere with the device's safe performance.

28:27.540 --> 28:33.380
So that within 60 days may represent an impossibility in certain situations.

28:33.380 --> 28:39.740
And in those cases, a manufacturer is going to need to undergo a full Part 806 with a recall.

28:50.370 --> 28:52.950
So I can't quote you a percentage.

28:52.950 --> 28:57.990
What I would say to you is the following from anecdotal from our experiences thus far.

28:58.130 --> 29:03.510
And that is that well, there are a few conditions here to talk about.

29:03.510 --> 29:09.390
Number one, the implantable devices represent the hardest fixes.

29:09.650 --> 29:27.970
So you want to think about the pacemakers, the ICDs, you know, the implanted defibrillators, neuromodulators, deep brain stimulators, things that are implanted are going to represent a much more complex fix and deployment.

29:28.670 --> 29:38.430
So now they don't represent a huge percentage as we were looking across the entire medical device portfolio of devices that we regulate.

29:38.430 --> 29:41.590
That's not a, you know, huge chunk.

29:41.630 --> 29:45.630
But you have that as one consideration.

29:45.630 --> 30:03.890
The other important consideration, which is relevant to this discussion also, is that we've seen a fair amount, particularly with these are high consequence types of vulnerabilities, what can be debate or disagreement across researchers and manufacturers,

30:03.890 --> 30:09.090
or researchers, manufacturers, and government with regard to what we like to call ground truth.

30:09.090 --> 30:14.790
Is this vulnerability, first of all, real in the sense of does this exist?

30:14.930 --> 30:19.130
Has there been an ability to validate its presence?

30:19.250 --> 30:40.360
And more importantly, or by taking it a step further, the impact of that vulnerability and the severity of that impact, where its capability of introducing patient harm that we call uncontrolled risk, or risk that we would consider to be not tolerable from the agency's perspective,

30:40.630 --> 30:51.910
not tolerable in the way it exists, and therefore requires manufacturer to do something with that device to reduce that risk to an acceptable level.

30:51.910 --> 31:06.910
So when you have disagreement across parties with regard to how to assess the assessment of that vulnerability and its impact, that already is going to prolong the timeframe out of the 60 days.

31:06.910 --> 31:18.690
So that's the other piece also that has resulted in some of the recalls that have been public, getting to that ground truth has been a very complicated process.

31:18.850 --> 31:31.370
And because it's well exceeded 60 days, it's been months in some cases, we've had a few that have gone a year and a half or longer, that has necessitated a recall.

31:31.370 --> 31:35.510
And so now we have draft pre-market guidance as an update.

31:35.510 --> 31:39.230
Can you tell us a little bit about that draft and what the next steps will be?

31:39.230 --> 31:39.670
Yeah.

31:40.130 --> 31:56.990
So one of the also important principles where we started out from is that with cyber security really rapidly evolving and knowing where we and the ecosystem were starting, we recognized that this is an area that we were going to need to iterate on, on a continuum.

31:56.990 --> 32:05.270
And then what we issued back a few years ago, that's fine for a foundation, as appropriate as a baseline kind of setting.

32:05.310 --> 32:23.870
But given everything that we have observed, everything that we have learned and that we've experienced over the past few years, whether it's WannaCry as an example, or NotPetya that was talked about earlier, whether it's vulnerabilities that have been brought to our attention by security researchers,

32:23.870 --> 32:35.930
we recognize through those learnings that, yeah, we got to be continuing to raise the bar and our expectations of manufacturers are that much more rigorous than they were in 2014.

32:35.930 --> 32:41.610
So 2014 sets the floor and the principles upon which we continue to build.

32:41.610 --> 32:42.700
Did you have a question?

33:03.440 --> 33:04.600
Yes.

33:04.600 --> 33:05.260
Yes.

33:05.260 --> 33:35.300
And those are examples of where we've had this protracted phase of trying to get to ground truth around vulnerabilities where we have had to insert ourselves as part of that process, working together with the security researchers, working with subject matter experts that are technically capable of being able to further understand exactly what the researchers have been able to demonstrate and working with the manufacturer.

33:35.820 --> 33:58.940
And that is a challenging space, that I'll take a detour for a second, is a reason why we put forward in 2018, in the spring, a medical device safety action plan for medical devices in general with an entire section on cybersecurity.

33:59.200 --> 34:06.760
And we put forward a proposal there for something that we call the CIMSAB, CyberMed Safety Analysis Board.

34:06.760 --> 34:27.380
And the concept behind the CyberMed Safety Analysis Board is just that, recognizing that we are challenged today, we have a gap of knowledge and understanding within the ecosystem where you can have vehement disagreement across different parties as to not just the vulnerability's existence,

34:27.380 --> 34:38.640
you know, I think that we can get to that point, but what is the impact of that vulnerability from a clinical perspective and therefore what action would need to be taken.

34:38.640 --> 34:59.840
So, having an entity that's more of a neutral body, a neutral entity that's a public-private model that includes technical experts, clinical experts, engineering experts, security experts, that government has a seat at the table as well, but government doesn't own this entity.

35:00.020 --> 35:26.180
Think somewhat like akin to an NTSB model, not exactly like that, but the idea, and not waiting for something bad to happen, but rather when a vulnerability of this high consequence is brought to our attention, it's, you know, we find ourselves having to reach out to multiple parties trying to get to this place of understanding really what the impact of that vulnerability is.

35:26.180 --> 35:48.220
How much better would it be if there was a pre-positioned body of experts that are able to come together in real time to fully assess through each kind of expert's lens, through different disciplines, really what the impact is of that vulnerability and what are the necessary next steps to take.

35:48.360 --> 35:53.440
So, that is something that we are proposing for initial seed funding.

35:53.440 --> 36:05.580
We're hoping that we can get some funding for that with the idea of it being subsequently owned and operated by the public-private sector, kind of through a public-private sector model.

36:06.300 --> 36:15.600
But going back to the new pre-market update, so in reflecting on what we've learned, we come to three basic principles.

36:15.600 --> 36:24.840
We call them trustworthiness, transparency, and resilience, recognizing that we've got to build those foundational principles into everything that we do.

36:24.840 --> 36:31.660
It has to start with the pre-market, but it really goes throughout the entire post-market lifespan of a device.

36:31.660 --> 36:40.720
However, you're only going to get to that if you build it from the start, from scratch, with all of those concepts in mind.

36:40.840 --> 36:44.900
And so, we released that in October.

36:44.900 --> 36:53.730
Some of the key elements there that were really important ahas for us, number one, was the concept of bill of materials, right?

36:53.730 --> 37:02.270
So, WannaCry had transpired and a recognition that you can only protect what you know that you have.

37:02.270 --> 37:24.210
For an end-user organization, for customers, for healthcare organizations that have thousands upon thousands, if not, in some cases, millions of medical devices residing on systems, how can they possibly protect their systems, protect the devices in terms of their performance if they do not know what the composition is,

37:24.210 --> 37:30.890
the components are of those devices, and whether there are vulnerabilities that have already been identified within those devices.

37:31.070 --> 37:47.130
So, while we are heavily engaged in a fantastic effort, a multi-stakeholder effort through NTIA, Department of Commerce, the SBOM, or the Software Transparency Initiative, which is fantastic.

37:47.430 --> 37:55.390
At the time, we also thought we were going to kind of even raise it a bar further and say, well, it's not just software, we're concerned about hardware as well.

37:55.530 --> 38:08.290
So, we called it the CBOM in the draft guidance, the Cybersecurity Bill of Materials, with the expectation that manufacturers actually would have to provide that to their customers, not just to us, to their customers.

38:08.410 --> 38:27.930
We've had a workshop since, our docket as well, the comments have come in, and there's been recognition across the board by people that are a lot more knowledgeable and experienced than us, that that's great, and that's aspirational, but that's maybe taking it down the road.

38:27.930 --> 38:34.550
Really, what you want to start with is, let's get that SBOM, let's get that Software Bill of Materials in place.

38:34.550 --> 38:54.510
So, as we're in the process right now of revising and kind of taking into account the comments that came into the guidance draft that we updated, the update that we issued in 2018, we're going to put that back as SBOM, as opposed to CBOM, and once we get there,

38:54.510 --> 39:00.090
then we'll move forward to CBOM as we continue to iterate and make things more rigorous.

39:00.090 --> 39:15.030
Some of the other pieces that were important there was putting in an expectation that manufacturers have to include threat modeling and documentation of threat modeling in the design of their devices.

39:15.250 --> 39:29.530
Third item that was really important was, manufacturers need to include in their pre-market submissions evidence that devices are capable of being updated and patched through the lifetime of their device.

39:29.530 --> 39:35.170
We are dealing with the pain of legacy devices right now.

39:35.490 --> 39:44.210
That has been an area that is going to be very difficult to address over the next few years.

39:44.210 --> 39:58.950
What we can do while we're trying to address that through one arm, certainly with new devices coming on the market, the expectation that they have greater resilience, that they are capable of being updated and patched.

39:58.950 --> 40:12.950
We want to see that as part of the pre-market submission, not to find out once a device is in use that this device is brittle and any attempt to trying to update it is actually going to knock over the device.

40:13.270 --> 40:19.110
So those are some of the concepts that we incorporated within the new pre-market guidance.

40:19.130 --> 40:31.310
And we're, again, once more in the process of revising it right now with the hopes of being able to release it again by the end of this calendar year.

40:32.070 --> 40:35.770
Okay, and to maximize time for questions, I will open it up.

40:35.770 --> 40:39.210
Are there any other questions that we can engage on?

40:42.610 --> 40:44.710
Could you use the mic?

40:44.730 --> 40:45.850
Thanks.

41:04.580 --> 41:06.240
Yeah, really good question.

41:06.240 --> 41:35.120
I think right now we are also dealing with the reality at FDA that we are responsible for making changes that were mandated by 21st Century Cures Act as well, which brings into account a lot of interesting questions around products that have some functions that are considered medical device functions,

41:35.120 --> 41:44.440
upon which we have direct authorities and some functions now which may not actually be considered medical devices.

41:44.440 --> 41:54.140
So you can imagine how sort of convoluted that can be as far as how to approach that from a cybersecurity perspective.

41:54.900 --> 42:03.520
Trying to create an artificial walling parts of it off from what we would look at versus what we wouldn't look at is very complex.

42:03.520 --> 42:17.820
We're working through some of those challenges now, but what I would say is the following, and I had this aha moment actually when I was sitting and listening to Alan Friedman the other day at B-Sides, you know, talk about the Software Transparency Initiative,

42:17.820 --> 42:23.640
the effort, and the group's work in terms of defining, well, what exactly is SBOM?

42:23.640 --> 42:24.700
What really is it?

42:24.700 --> 42:29.520
And he said where the group had come to was X includes Y.

42:29.520 --> 42:35.880
Let's not talk about functions right now or, you know, let's talk about what's included within that.

42:35.880 --> 42:47.260
And for me, that really resonated because that meant, you know, the SBOM is really intended to be that protective measure, that ability to mitigate for the end user, for the customer.

42:47.540 --> 42:57.140
And so trying to separate out or divorce out pieces of it that would not necessarily be under our authority doesn't make sense.

42:57.140 --> 43:15.980
As we look towards SBOM, we would be looking at the product in its totality, even those components of the product that may not necessarily have a direct medical device performance function, but that wouldn't matter to the end user, right?

43:15.980 --> 43:19.440
Their concern is for the vulnerability and their exposure.

43:19.460 --> 43:24.820
So that's, you know, that's kind of how we're thinking about it right now.

43:29.160 --> 43:35.080
I was curious about the scope of the vulnerabilities that your team is responsible for addressing.

43:35.080 --> 43:43.420
Is it just physical hazards or the function of the hardware device or do you also look at things like user privacy, for example, or other?

43:43.440 --> 43:45.460
Yeah, that's a really great question.

43:45.460 --> 43:46.820
I didn't cover that at all.

43:46.820 --> 43:59.340
So the interesting thing is that across the Department of Health and Human Services, different agencies, different operating divisions have different authorities that they are responsible for.

44:00.100 --> 44:09.500
ONC, I'm sorry, not ONC, OCR, Office of Civil Rights within HHS, is responsible for HIPAA privacy and security.

44:09.500 --> 44:12.240
That's not under our remit.

44:12.380 --> 44:23.320
And so while we would not minimize the importance of privacy, of patient information, clearly that's very important.

44:23.320 --> 44:38.780
Our authorities and our mission is specific to the safe and effective performance of devices and protecting patients from physical harm if the device was not performing properly, it was not performing at all.

44:38.780 --> 44:55.640
So what we did in guidance that we've previously issued, though, is we state very clearly, manufacturers, you are still under other obligations that you need to abide by, including HIPAA privacy and security.

44:55.640 --> 44:59.800
Here's what we're doing to make it easier for you to abide by that.

44:59.800 --> 45:09.420
So we have something called controlled vulnerabilities, vulnerabilities that in their assessment don't necessarily result in a patient harm.

45:09.420 --> 45:21.000
What we've said there is that changes that are made solely to strengthen cybersecurity and data privacy-related issues would fall in that category.

45:21.000 --> 45:25.560
They are considered, from our perspective, routine updates and patches.

45:25.560 --> 45:29.020
They do not come to the agency for re-review.

45:29.020 --> 45:31.200
Manufacturers, you should just go and do that.

45:31.200 --> 45:44.660
You should just take those proactive steps to make sure that you can then be in compliance with the other regulations under other parts of government that you are required to abide by.

45:44.660 --> 45:47.360
And we're not going to impede that process.

45:47.360 --> 45:50.100
We're not going to delay that because we say we need to see those.

45:50.100 --> 45:51.940
We do not need to see those.

45:55.420 --> 45:56.920
Other questions?

46:00.560 --> 46:07.420
What do you foresee to be the biggest challenge going forward in the next five years for the FDA to the point of security?

46:08.360 --> 46:09.800
Biggest challenge?

46:10.340 --> 46:12.160
There's a few of them.

46:12.220 --> 46:14.000
Some of them are wrapped together.

46:14.000 --> 46:31.700
I think holistically across the ecosystem, so it's not just FDA's challenge and this is where we really have to bring other parties together, it is the legacy system and the legacy device issue.

46:31.700 --> 47:02.940
And I mentioned that because of the economics associated with that as well and the recognition that while new devices can be made available to healthcare organizations, healthcare organizations are not necessarily well-poised financially or through their business models to be able to take in those new devices in order to get rid of the old legacy drag.

47:03.700 --> 47:31.520
And so you're left with certain systems that are not capable of being updated and patched and you're left therefore with certain exposures and vulnerabilities and recognizing that a cyber campaign can occur at any time and can exploit these vulnerabilities that leaves our critical infrastructure vulnerable.

47:31.520 --> 47:36.360
So how do we address that from multiple facets?

47:36.360 --> 47:41.400
Recognizing that, yeah, we have some tools in our back pocket that we could pull out.

47:41.400 --> 47:45.100
Some of them are a little bit drastic that we would rather not move to.

47:45.100 --> 48:06.680
We'd like to figure out through efforts of the public and private sector working together and there are a lot of different consortia doing that under even what's called the big umbrella, the Healthcare Sector Coordinating Council to really analyze some of the economics here and then what are the options,

48:06.680 --> 48:18.480
what might be the incentives that could be put forward to get these really older devices out of use and bring newer devices that are more rigorous and more secure in place.

48:19.100 --> 48:35.980
I don't know, Seth, if there's anything that you want to add, but that's, you know, we've got lots and lots of challenges of different scale, but when I think of something that's really big scale that's going to require a whole universe of effort, that's probably one of the biggest ones.

48:36.680 --> 48:39.860
And with that, we will have to wrap things up.

48:39.860 --> 48:41.440
Let's all thank Dr.

48:41.440 --> 48:47.220
Suzanne Schwartz for joining us today and let me thank you by encouraging you to take a sticker on your way out.

48:47.580 --> 48:48.340
Thank you.

48:48.340 --> 48:49.020
Thank you.
