0	4560	Hi everyone. Thank you so much for coming to my talk.
5140	10280	If you'd like to... I have some detailed slides, so if you'd like to get into those details,
10280	15380	please feel free to click the QR code. I promise it's safe. I don't know how to make it not safe.
15700	19380	So... and follow along for your use.
19960	24960	My name is Lacey Harbour. I have a background in molecular biology, but fell
24960	30060	unfortunately or fortunately into regulatory and quality and clinical studies.
30060	34860	And today, I want to present some ideas to everyone here
34860	41920	about how medical devices are being made and how they're going to be made in the future.
41920	46360	And during this process, I'm going to highlight some areas of concern
46360	50280	that is particularly applicable to our cybersecurity partners.
50280	55660	And how we're not really ready for the future of the medical device
55660	60420	world. So, it goes without saying that
60420	65240	AI is changing everything. AI is particularly changing
65240	70640	the healthcare ecosystem. And this is a good thing for
70640	75600	us because us as a population, we're aging. And our population
75600	80620	isn't growing with new members as much as it used to. With that,
80620	85760	we're going to need a new point of care, new care, standard of care
85760	91260	treatments. And that's not going to be doable without AI.
91660	96040	So, AI being integrated into the products
96040	100600	are changing the scope of these products. And my world is regulatory.
100600	105400	We like to bucket products into regulated and unregulated product. You can have the exact
105400	110500	same instrument or device or whatever. And if you define it to be intended for
110640	114960	a medical purpose like diagnosis, it's a medical device. So, you have all these controls in place
114960	120020	to deal with that product, to develop it, to manage it, to monitor it.
120560	125380	But if it's not a medical device, then you don't have to monitor it to the same level
125380	129780	of scrutiny. And that can be a concern.
131880	135520	So, when something's called a medical
135520	140520	device, particularly in the U.S., it's all depending on its intended use.
141080	145480	Now, with the ecosystem changing, with the scope of the products
145480	150640	changing, the intended use is not just the only thing we need to be worried about or be mindful
150640	155980	of. We need to be mindful of how it's being used in the health care system,
155980	160900	what data sources and attributes it's plugging into. Who are the
160900	165640	end users? It's not just the physician anymore. There's different voice of customers that need to be paid
165640	170920	attention to, especially when you're building a product that should be secure
170920	175700	and safe and effective across the entire use
175700	181320	and its entire ecosystem. In the past, I used to have to be worried
181320	186300	just about the FDA in the U.S. market. Now, I have to be mindful of
186300	191520	all of these entities, these regulatory entities that are listed on the screen.
191620	196280	And that's just the U.S. market. I'm part of an FDA collaborative
196280	200940	community. I'm one of the co-leaders of it. And we have an NLP that just
200940	205780	scans the Internet, picking up sources that we've deemed
205780	211100	reliable and allowing us to analyze all these global sources into a one-stop
211100	215700	shop. Within this year alone, there have been 85
215700	221060	sources or materials that we as a regulatory equality partners
221060	226100	must analyze and communicate to our companies to take action upon.
226100	231340	And I don't know if you really appreciate how much freaking work that is and how damn near impossible
231340	236020	that is. Because not only do I have to analyze it, and then I have to teach the executives
236020	241400	this information, then I have to get buy-in that they need to update procedures and policies.
241420	245980	We have to do a business risk impact and see how much, you know, it just keeps going and going and going.
245980	251120	Ultimately, you have to train the people within the manufacturing line even what they can
251120	256240	and cannot do. This is a huge task. And of course, it's going to be easier based on the
256240	261260	size of your company or maybe not. Each company has their own problems
261260	266580	with this. It's not just the regulatory ecosystem.
266580	272580	It's also the relationship with our customers. So as medical device manufacturers,
272580	275920	you know, you always think the customer is the physician or the patient.
275920	281060	But that's not always the case. We are plugging in our products into health delivery
281060	285740	organizations. And so they have to navigate their own set of regulations.
285860	290820	And they have their requirements. And so we have to make sure that our products interact
290820	296040	with that. Also, with the future of these products, again, AI enablement
296040	300900	is a really data-hungry process no matter where it's plugged into. And
300900	306000	who owns the data? Well, the EHR companies like Epic. You think they want to
306000	310960	share that data with the medical device manufacturers so that we can have good quality outputs?
310960	316100	Not necessarily. So we have to build those relationships with the medical device manufacturers,
316100	321000	the EMR, and any other applications that are plugging in to the hospital
321000	326260	delivery organization or HDOs. So FDA
326260	331240	is seeing this. And I like to... they're not the only
331240	336260	agency globally that's really making progress. But they are really being public about
336260	340960	it. And it's accessible. Any of you guys can go on the FDA website and see all sorts of
340960	346180	initiatives that they're putting into place to make sure that the public is being heard.
346180	350560	And they're actually engaging patients even. They're really paying attention to the voice of customer
350560	354860	and how these products are truly used. For example,
354860	361020	they'll ask about the workflow in a laboratory even.
361520	366520	And so I highly recommend and encourage everyone to go check out these resources.
366740	370720	But when the FDA is in particular... and they're working with an international
370720	375460	group called the IMDRF. When the FDA is hearing
375460	380820	and seeing where the industry is going, they're
380820	385720	trying to grow along with industry so that they're not holding us
385720	390600	back. And one particular use case I'm going to talk to you guys about today
390600	395620	is called the medical device production system. But before I can get
395620	401240	into this future device, essentially, I want to level set everybody.
402380	406180	Who's here familiar with the total product lifecycle?
406980	412160	I have a handful of people. All right. So the total product lifecycle
412160	417860	is basically the inception of a product all the way to the end of life of a product.
417940	422300	And so I put together here a representation
422300	427880	of any kind of QMS or quality management system that represents a total product lifecycle.
427880	432780	And it looks like a cell. And I really... maybe the biologist is being a little biased
432780	437520	in me, but it really is like feedback loops, like what you get in a cellular function.
437520	442680	So you could break it down, in my mind, a total product lifecycle into design
442680	447540	controls, production and process controls, and then there's feedback loops for improvements. And any of those
447540	452360	improvements could be like failure on the field. It could be a non-conformance from a manufacturing line.
452360	457100	It could be a corrective action that's discovered from anything else or regulatory updates.
457640	462600	But this is too complicated to really talk about and use.
462600	467280	So I further simplified the total product lifecycle schema
467280	472420	into these two basic buckets. So you can see where I have the
472420	477460	QMS and then I've broken it down to design controls and production and
477460	483140	process controls, where you have a prototype of a device. It can be a digital device
483140	487580	or a cyber device, as they call it in the US. It could be
488520	491740	a tongue depressor, depending on the classification.
492300	498020	But say that you have the prototype. You test it.
498020	503160	Hey, it's good. I'm going to now scale up and manufacture this.
503160	508680	So then you transfer to manufacturing. You test it. Hey, the manufacturing is working well.
508680	513140	Now we can actually start mass-producing these products. And then you
513140	518260	manufacture and then you test it again. And then finally, if they match all the quality control
518260	523560	or QC checks, that device is released out into the wild or software or whatever.
523980	528260	Within each of these steps, there's feedback loops. So something bad
528260	533400	can happen during this process to kick you all the way back to design of, you know, feasibility
533400	538200	and design phase. But at the end of the day, the take-home message is
538200	543180	the output, the final device, is from that whole end-to-end total
543180	545420	product lifecycle within the QMS.
548540	553320	Now that I've kind of given you a very high level of the basics of
553320	558620	what currently exists, let's talk about the future. The future of our products
558620	563740	is going to be precise. It's going to be personalized.
563740	568740	It's going to be at the point of care. And there are several different industries
568740	573300	and sectors and technologies that are helping pushing our
573300	577320	standard of care practice to this point. But
577320	583100	the orthopedic sector has really been
583100	587800	leveraging the 3D printing technology to bring us closer
587800	592420	to that precise, personalized, and point-of-care
592420	597380	product. And I'm going to reintroduce the idea of a medical device
597380	602260	production system to you. And so this isn't the only use case
602260	607320	involved that does this. But these orthopedics, I guess
607320	612380	because, I mean, to be honest, it's human carpentry. So the complexity isn't as
612380	617420	much as maybe other devices. So they're moving at a very fast pace. And the FDA
617420	622400	saw that, and so the IMDRF. So they've started to put together some loose guidelines to help
622400	626040	lead this medical device production system.
626840	631380	So what is the medical device production system, or MDPS?
631380	636080	So let's go back again to the schema I showed you all before. Again,
636080	640980	the output is a device, right? It's pretty simple, of a total product
640980	646000	lifecycle of the QMS. Well, a medical production system turns
646000	651100	that on its head. Now, the output, the device, isn't
651100	656400	the final component. And this is a little hard to wrap your head around. It includes
656400	661460	the manufacturing. It includes the design phase. So what happens is
661460	666860	I'm a medical device manufacturer, and I'm going to make a medical device production system, an MDPS.
666860	672060	I'm going to design the production line just as I would before.
672180	676240	Except now, I'm going to realize that this production
676240	681440	system is going to be part of the device in the healthcare delivery organization.
681440	685940	So there will be manufacturing in the hospital. And
685940	691400	then I will allow a certain range of scope or specs that
691400	696540	can be allowed for the hospital to manufacture. And so from that end-to-end process,
696540	701260	they can design the product, and they can manufacture it, and then the output is
701260	706120	part of the entire device. The output alone is no longer the singular device.
706560	711420	So if any of you are familiar with the quality system
711420	716300	regulation or 13485 standards, your wheels may be turning like, how the
716300	720320	hell do you control this? And that is a big question. But
721560	726380	the precursors to this are existing right now. Here I have
726380	731600	an image of some 3D-printed EBM joint implants
732140	736700	made. They're called trabecular titanium, and they allow for bone engrowth.
736700	741680	And they are all patient-matched. And these are printed in the hospital
741680	746700	as we speak from a medical device manufacturer. And so
746700	751080	I'm going to kind of introduce at a high level this current
751080	756280	process, and then we're going to talk about how this is going to become
756280	761220	the medical device production system. So say I'm
761340	766100	a patient, I have a really janked up shoulder, and the physician
766100	771120	says, hey, the standard products on the market, they're not going to work for you. We need to
771120	776200	give you a custom device. So they're going to send me to get a fresh CT scan. I'm going to
776200	781780	get my first CT scan, and that DICOM file is going to go to the radiologist, to the surgeons,
781780	786060	and they're going to segment that file with the medical device
786060	791960	manufacturer or the designer. They will identify the defects,
791960	796380	and then they will agree on how the physician wants to approach it.
796380	801020	So the physician is really close with the designer through this whole process.
802220	806280	After they discuss it, the designer will then take
806280	811160	that DICOM file and convert it to an STL file within a 3D CAD program
811160	815760	like Magix or something similar to that. And they will
817200	821200	propose potential designs. And the designs don't come out of anywhere. It's based
821200	826060	off of historical data, recall data from standard devices, but it will be
826060	831800	matching how the physician wants to perform the surgery and what the physician actually wants to do.
831860	836660	And so there's a feedback loop there until the physician gets the green light, and then it is released.
837720	841380	But just because you can design something doesn't mean you can
841380	846780	actually print it. Within a 3D printer, there's XYZ zones,
846780	851200	and each zone has a different behavior. They're their own animal.
851840	856200	So what happens is that the designer will give the STL file
856200	861240	to the production engineer, and then they will convert it to a machine file and determine
861240	866320	if it's possible to actually 3D print that geometry. And some of that may be tied to
866320	871420	demand planning because you obviously, you have a large bed, and
871420	875140	you may be printing more than one patient's devices at a time. And so
876040	880860	you may have to strategize that. But it's not uncommon for that device to be
880860	885400	kicked back to the physician, essentially, when the production engineer says,
885400	890740	no, we can't do that. After it's printed, there are a lot of supports
890740	895680	that need to be removed. And so typically it's like a CNC or a saw that helps remove
895680	900700	those products, and then it goes to a QC check. And the QC check is just not the
900700	904920	outside dimension, but the inside integrity. These devices have to
905840	910780	resist weight and movement. And so if there's any kind of flaw internally,
910780	915540	that can cause device failure. So there's a lot of checks that have to happen with QC
915540	920740	samples. So what might be of interest is, let's
920740	925700	dissect a little bit more detail of the first half of that from
925700	930220	the relationship of the health delivery organization and the medical device
930220	935800	manufacturer. I'm not going to go into all the detail of every
935800	940100	bit of this schema, but this is just a representation of what that
940900	945400	data interaction might look like. But what I really want to highlight, and I see
945400	950320	some of you shaking your head or maybe have some concerns, you might have noticed that there's
950320	955900	something called a PAC server, and that holds all of the DICOM data for pretty much everything.
956300	960500	And how does that data get to the medical device manufacturer will depend on the relationship
960500	965300	the medical device manufacturer has with the health delivery organization. It could be a nurse
965300	970520	downloading it, it could be on a flash drive and handing it to the device manufacturer,
970520	975500	it could be an email, it could be a CD, or it could be connected directly
975500	980740	into the medical device's server. The medical device
980740	985380	manufacturer may have a software that sits on top of the PAC server, or
985380	990460	they may have a relationship where they have access to the EMR. And now
990460	995600	if they have access to the EMR, they have to have a relationship with the EMR. So the problem
995600	1000940	is that I want to highlight is, I just want to make a note
1000940	1006800	that medical device manufacturers are not HIPAA covered entities.
1006800	1011120	And I'm just going to leave that there and let you kind of think on what that means.
1011520	1016300	So once the design
1016300	1021260	is considered safe, at least from a
1021560	1025160	failure mode analysis perspective, then it goes back to
1025160	1030160	the internally. So now we're completely inside the medical
1030160	1033800	device manufacturer's side of the network.
1034880	1040540	Many device manufacturers are moving forward to using what's called machine learning operations.
1040620	1045200	And this is just a high level example of how that
1045200	1050260	can look within the network. You have containers, of course, that are cross-talking with different
1051860	1055900	instruments. And then you can also have these containers cross-talking
1055900	1060720	with other systems, like your product lifecycle management system,
1060720	1065480	your quality management system, or your ERPs,
1065780	1070740	or even other sites, if you have multiple sites interacting. And that's something that would be a goal
1070740	1075740	for medical device manufacturers with MDPS. So it'd be nice to have communications to see the success
1075740	1080420	of the products across all of your deployed MDPSs.
1081460	1086640	And so you can see here that there, again,
1086640	1091440	you may be the design engineer. How did you get the data from the designer?
1091440	1095500	Well, it could have been put up on a server. It could have been a flash drive handed to you.
1095500	1100620	And you work on this product for the machine file, the machine planning, on your
1100620	1105680	personal laptop or on your work laptop, because the data processing for this is
1105680	1111340	quite substantial. And then how does that data get to the EBM in the first place?
1111340	1116020	Again, it could be a flash drive. There's a lot of flash drives used in manufacturing,
1116020	1120620	especially with smaller firms. So there's
1121220	1126200	concerns there. But this isn't MDPS.
1126200	1131120	This whole workflow that I presented, it's precursor to MDPS. And it exists
1131120	1136400	right now. Here are just some firms that are doing this right now. And this isn't even all of them.
1136580	1140320	There's a lot of research going into how to get the MDPS.
1142180	1146540	So what is the MDPS? Well, just for simplicity
1146540	1151040	purposes, think about that whole workflow and put it in a box.
1151040	1156200	You get the CT scan, you plug that DICOM in this box. There might be
1156200	1161040	people remote, like designers or even people monitoring the success of the
1161040	1165860	instrument remotely. And the physician is
1165860	1171000	working with some kind of software interface to say, hey, I want the design to look this
1171000	1176060	way. And then they press go and out pops your final device.
1176060	1181120	Now, if anyone's familiar with 3D printing, the post-processing is going to be a problem for the
1181120	1185280	future MDPS, but it's not going to be unattainable.
1185760	1190620	This is, we are working, technology is being researched, mineralization is being researched.
1190620	1194100	This is not like it's going to be 50 years from now.
1195440	1200100	What's going to make this even more powerful is that you're going to have feedback from the surgical suite
1200100	1205360	into the MDPS. So you're going to get clinical outcomes. But again, to get that, you have to
1205360	1208400	have the relationship with the medical device manufacturer.
1210600	1215420	So AI is going to be necessary to do this. There's no way
1215420	1220480	that you're not going to be able to have this without. So let's highlight some of the areas where AI
1221200	1225240	is being researched even right now and where it's going to be plugged into this MDPS.
1225240	1231460	I mean, if you look this up online, you're going to see a slew of papers because this is the low-hanging fruit.
1231460	1234400	Again, human carpentry, everyone's going for this.
1234880	1240620	First place, automatic segmentation. Have a mask, highlight the defect,
1240620	1245720	give you recommendations on the defect. The radiologist, the surgeon, or whomever can look at that
1246320	1249780	and give yes or no feedback. The next place
1252080	1254560	is the design.
1254560	1258700	You can have all of your data from your field feedback
1259660	1264260	from the designs that exist that are standard to automatically generate design.
1264260	1269340	And all the designer has to do is say go, no go, and the physician
1269340	1274460	have their inputs as well. And you can use LLMs to generate the surgical technique,
1274460	1279320	which is also considered part of the labeling, which is part of the design or the device.
1279880	1285360	Finally, the demand planning, the design transfer.
1285740	1289860	You can have multiple products in the queue
1289860	1295700	and you can determine, hey, this device needs to be manufactured in this zone,
1295700	1301160	but this patient is suicidal, so they're a higher risk. So that would adjust the demand planning.
1301160	1305900	And that's something, again, you need good connection with the physician and the health
1305900	1311300	HCPs to determine what's correct. And then, of course, the enabling of manufacturing
1311300	1316580	with ML is already in progress and they're just going to miniaturize that.
1316580	1321340	And it would make everything more visible and you can monitor the
1321340	1326360	manufacturing in real time. So I can't
1326360	1331340	stress enough that maybe with the high level schemas that I presented to everybody, you might
1331340	1336420	kind of detect that we, again, are not ready for this, but the technology is getting
1336420	1340640	there and we need it. We need it now. We need it yesterday.
1340960	1346360	And so medical device manufacturers need to start being aware of the
1346360	1350620	three customers down the road so that we can have safe and effective
1350620	1355660	products that are consented, that are safe, that have
1355660	1360700	encrypted or private. And that's not something medical device manufacturers are always
1360700	1366080	mindful of. We are really in a next-gen humanity
1366080	1370700	moment here. Here's an example of 3D printing
1370700	1376060	autologous bone. So you have a defect. Instead of a joint implant,
1376060	1381040	you 3D print a mold that's based off the defect and then you use a bioreactor
1381040	1386040	to print your bone. It's not just orthopedics. It's just the low-hanging fruit. I mean,
1386040	1390080	the application is huge and the MDPS is going to be amazing.
1390500	1396000	So we are in a deus ex-human moment. This is something I
1396000	1401080	stole from my friend Nina. And it's true. We are making new humans.
1401080	1406200	And so how are we going to do this safe? I encourage everyone to go to the resources I pointed
1406200	1410300	out. Kind of look at the schema. What can you do to encourage this and where we can all
1410900	1415900	have a better future and better health care practice and better standard of care.
1415960	1418840	So I appreciate everyone. Thank you.